FDA Adverse Event Malfunction Summary report: N

ECHELON 60

MDR report key: 1970891 · Received January 26, 2011

Report

Report Number
3005075853-2011-00306
Event Type
Malfunction
Date Received
January 26, 2011
Date of Event
December 30, 2010
Report Date
December 30, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K051002
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT ONE EC60 DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO CARTRIDGE RELOAD PRESENT. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE 3-STROKE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. EVENT COULD NOT BE CONFIRMED AS NO CARTRIDGE WAS RECEIVED FOR ANALYSIS. THE ANALYSIS RESULTS SHOWED THAT ONE ECR60G CARTRIDGE RELOAD WAS RECEIVED IN GOOD VISUAL CONDITION AND FULLY FIRED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOW ANTERIOR RESECTION, THE SURGEON COULD NOT FIRE THE ECHELON SMOOTHLY AND DID NOT GET GOOD FORMATION ON THE COLON. IT WAS NOT REPORTED HOW THE PROCEDURE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.