PLUER-EVAC ADULT-PED DRY/WET SYSTEM
Report
- Report Number
- 3004365956-2011-00011
- Event Type
- Malfunction
- Date Received
- January 4, 2011
- Date of Event
- December 9, 2010
- Report Date
- December 9, 2010
- Manufacturer
- TELEFLEX MEDICAL
- Product Code
- KDQ
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- NURSE
Narratives
EVALUATION METHOD: DEVICE HISTORY RECORD (DHR) REVIEW AND COMPLAINT HISTORY REVIEW. RESULTS: DHR REVIEW FOR THE REPORTED LOT NUMBER SHOWED NO ISSUES OR DISCREPANCIES WHICH COULD POTENTIALLY RELATE TO THE COMPLAINT REPORTED. NO NON-CONFORMANCE REPORTS WERE ORIGINATED FOR THE LOT IN QUESTION THAT CAN BE ASSOCIATED TO THE COMPLAINT REPORTED. THE DHR SHOWS THAT THE PRODUCT WAS ASSEMBLED AND INSPECTED ACCORDING TO SPECIFICATIONS. COMPLAINT HISTORY REVIEW FOR ONE YEAR WAS CONDUCTED ON THE CATALOG NUMBER IN QUESTION FROM (B)(4) 2009 TO (B)(4) 2010. TWO ADDITIONAL COMPLAINTS WERE RECEIVED FOR THIS CATALOG NUMBER. A ONE YEAR COMPLAINT HISTORY REVIEW ON THE PRODUCT FAMILY WAS CONDUCTED FROM (B)(4) 2009 TO (B)(4) 2010. ONE ADDITIONAL COMPLAINT WAS REPORTED. AN INVESTIGATION WAS PERFORMED REVIEWING THE ENTIRE PROCESS STARTING FROM THE MATERIAL RECEIVED AND UNTIL IT DEPARTS TO THE STERILIZER AND IT WAS FOUND THAT THE PROPER CONTROLS ARE ESTABLISHED IN ORDER TO DETECT IF THE ORANGE FLOAT IS PRESENT AND WORKING ACCORDING TO SPECIFICATIONS. THIS TEST IS PERFORMED AT 100% ON ALL UNITS AS PER PROTOCOL. VERIFICATION IS ESTABLISHED ON QUALITY INSTRUCTIONS (FLOAT ELEVATION AND DROP TEST). CONCLUSIONS: NO EVALUATION WILL BE PERFORMED. NO CONCLUSION CAN BE DRAWN. CUSTOMER COMPLAINT CANNOT BE CONFIRMED DUE TO LACK OF SAMPLE FOR INVESTIGATION. NO CORRECTIVE ACTION CAN BE ESTABLISHED AT THIS TIME.
THE EVENT IS REPORTED AS: THE FLOAT DOES NOT APPEAR IN THE INDICATOR WINDOW AT MAXIMUM SUCTION CAPACITY FROM THE WALL UNIT. THE UNIT WAS REPLACED WITHOUT A PROBLEM. NO PATIENT INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PLUER-EVAC ADULT-PED DRY/WET SYSTEM | PLUER-EVAC CHEST DRAINAGE SYSTEM | KDQ | TELEFLEX MEDICAL | NA | 02K1000181 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |