FDA Adverse Event Malfunction Summary report: N

PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED

MDR report key: 1970817 · Received January 25, 2011

Report

Report Number
1423500-2011-01039
Event Type
Malfunction
Date Received
January 25, 2011
Date of Event
December 6, 2010
Report Date
January 4, 2011
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS IS REPORT 1 OF 2 ASSOCIATED WITH THIS DEVICE.THE DEVICE HAS BEEN RECEIVED AT BAXTER FOR EVALUATION. HOWEVER, THE EVALUATION IS NOT COMPLETE AT THIS TIME. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). THE HOMECHOICE (HC) DEVICE WAS RECEIVED FOR EVALUATION AT THE PRODUCT ANALYSIS LAB (PAL). A REVIEW OF THE DEVICE LOGS CONFIRMED THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT. EXTERNAL & INTERNAL VISUAL INSPECTIONS REVEALED NO PROBLEMS. THE PAL EVALUATED THE DEVICE AND NO FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE IIPV FOUND IN THE DEVICE LOGS. THE DEVICE FUNCTIONED NORMALLY DURING PAL TESTING. THE CAUSE OF THE PEDIATRIC IIPV WAS DETERMINED TO BE INSUFFICIENT DRAIN DUE TO ONE OR MORE CYCLES ADVANCING TO THE NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. A SERVICE HISTORY REVIEW REVEALED THAT NO ISSUES THAT MAY HAVE CONTRIBUTED TO THE PROBLEM OF IIPV. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE (HC) MACHINE, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) WAS IDENTIFIED ON (B)(6) 2010 DURING DRAIN CYCLE 3. THE PATIENT'S ULTRAFILTRATION (UF) READING WAS 154 ML, INDICATING THE HOME PATIENT (HP) DRAINED 154 ML MORE THAN THE LARGEST PRESCRIBED FILL VOLUME OF 500 ML. THIS MEANT THE TOTAL DRAIN VOLUME WAS 654 ML, WHICH MEETS THE IIPV CRITERIA. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. DURING A FOLLOW UP WITH THE NURSE, SHE WAS NOTIFIED OF THE INCIDENT OF IIPV THAT WAS FOUND DURING THE DEVICE LOG REVIEW.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 5 YR