PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED
Report
- Report Number
- 1423500-2011-01039
- Event Type
- Malfunction
- Date Received
- January 25, 2011
- Date of Event
- December 6, 2010
- Report Date
- January 4, 2011
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K053512
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS IS REPORT 1 OF 2 ASSOCIATED WITH THIS DEVICE.THE DEVICE HAS BEEN RECEIVED AT BAXTER FOR EVALUATION. HOWEVER, THE EVALUATION IS NOT COMPLETE AT THIS TIME. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
(B)(4). THE HOMECHOICE (HC) DEVICE WAS RECEIVED FOR EVALUATION AT THE PRODUCT ANALYSIS LAB (PAL). A REVIEW OF THE DEVICE LOGS CONFIRMED THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT. EXTERNAL & INTERNAL VISUAL INSPECTIONS REVEALED NO PROBLEMS. THE PAL EVALUATED THE DEVICE AND NO FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE IIPV FOUND IN THE DEVICE LOGS. THE DEVICE FUNCTIONED NORMALLY DURING PAL TESTING. THE CAUSE OF THE PEDIATRIC IIPV WAS DETERMINED TO BE INSUFFICIENT DRAIN DUE TO ONE OR MORE CYCLES ADVANCING TO THE NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. A SERVICE HISTORY REVIEW REVEALED THAT NO ISSUES THAT MAY HAVE CONTRIBUTED TO THE PROBLEM OF IIPV. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
DURING EVALUATION OF A RETURNED HOMECHOICE (HC) MACHINE, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) WAS IDENTIFIED ON (B)(6) 2010 DURING DRAIN CYCLE 3. THE PATIENT'S ULTRAFILTRATION (UF) READING WAS 154 ML, INDICATING THE HOME PATIENT (HP) DRAINED 154 ML MORE THAN THE LARGEST PRESCRIBED FILL VOLUME OF 500 ML. THIS MEANT THE TOTAL DRAIN VOLUME WAS 654 ML, WHICH MEETS THE IIPV CRITERIA. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. DURING A FOLLOW UP WITH THE NURSE, SHE WAS NOTIFIED OF THE INCIDENT OF IIPV THAT WAS FOUND DURING THE DEVICE LOG REVIEW.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 5 YR |