FDA Adverse Event Malfunction Summary report: N

IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM

MDR report key: 19706985 · Received July 9, 2024

Report

Report Number
0002023141-2024-02303
Event Type
Malfunction
Date Received
July 9, 2024
Date of Event
March 12, 2024
Report Date
November 25, 2024
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019508
PMA / PMN Number
K061410
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). A4: PATIENT WEIGHT UNKNOWN / NOT PROVIDED D4: ADDITIONAL DEVICE INFORMATION UNKNOWN / NOT PROVIDED G4: PREMARKET IDENTIFICATION K061410/K011028/K013227 H4: DEVICE MANUFACTURER DATE UNKNOWN / NOT PROVIDED

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). THE FOLLOWING FIELDS HAVE BEEN UPDATED: B3. DATE OF EVENT B4: DATE OF THIS REPORT B5: DESCRIBE EVENT OR PROBLEM D4: ADDITIONAL DEVICE INFORMATION D9: DEVICE AVAILABILITY G3: DATE RECEIVED BY MANUFACTURER G6: TYPE OF REPORT H1: TYPE OF REPORTABLE EVENT H2: FOLLOW UP TYPE H3: DEVICE EVALUATED BY MANUFACTURER H4: DEVICE MANUFACTURER DATE H6: ADVERSE EVENT PROBLEM H10: ADDITIONAL NARRATIVE ZIMVIE RECEIVED ONE (1) TSVB10, (IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM)FOR EVALUATION. VISUAL EVALUATION / FUNCTIONAL TEST WAS PERFORMED, IMPLANT AND MOUNT ARE STUCK TOGETHER. UNABLE TO REMOVE MOUNT. DOES NOT DISENGAGE. MALFUNCTION. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1273452. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1273452 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: DENTAL : FUNCTIONAL : DOES NOT DISENGAGE/RELEASE BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WAS TORQUE APPLIED DURING PLACEMENT/SEATING EXCEEDS RECOMMENDED VALUE. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION DID OCCUR. THE IMPLANT AND MOUNT ARE STUCK TOGETHER. UNABLE TO REMOVE MOUNT. THE REPORTED EVENT IS CONFIRMED. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCTS WERE NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.

Description of Event or Problem · 0

THE DOCTOR REPORTS THAT THE MOUNT IS STUCK TO THE IMPLANT. AFFECTED DENTAL POSITION = 13. BONE TYPE = III. PROCEDURE COMPLETED WITH ANOTHER IMPLANT.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
545900 IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM DENTAL IMPLANT DZE ZIMMER DENTAL 1273452 00889024019508

Patients

Seq Age Sex Outcome Treatment
1 64 YR Female