FDA Adverse Event Malfunction Summary report: N

PUMP,COLLEAGUE CXE SINGLECHANNEL COLOUR-LOANER

MDR report key: 1970674 · Received January 25, 2011

Report

Report Number
6000001-2011-00443
Event Type
Malfunction
Date Received
January 25, 2011
Date of Event
July 7, 2010
Report Date
November 9, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Removal / Correction Number
6000001-12/1/08-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE CAUSE WAS DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND HARNESS HAVE BEEN REPLACED. A TREND REVIEW HAS BEEN PERFORMED AND THERE HAVE BEEN SIMILAR REPORTS MADE TO BAXTER. THE ROOT CAUSE WILL CONTINUE TO BE INVESTIGATED THROUGH CAPA (CORRECTIVE AND PREVENTIVE ACTION) INVESTIGATION (B)(4). THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS 6.13.92, CATEGORIZED AS REMEDIATED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED BAXTER (B)(4) TO REPORT A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY. DURING THE PRODUCT EVALUATION, IT WAS DISCOVERED THAT A BATTERY DEPLETED SET ALARM OCCURRED DURING INFUSION, WHICH CAUSED AN INTERRUPTION DURING DELIVERY. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS CURRENTLY UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP,COLLEAGUE CXE SINGLECHANNEL COLOUR-LOANER PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1