FDA Adverse Event Malfunction Summary report: N

AUTOMATED PD SET W/CASSETTE4 PRONG

MDR report key: 1970570 · Received January 25, 2011

Report

Report Number
1423500-2011-01005
Event Type
Malfunction
Date Received
January 25, 2011
Date of Event
January 1, 2011
Report Date
January 1, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE; HOWEVER, PER THE COMPLAINT INFORMATION THE CAUSE OF THE SE 2240 IS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE AFTER THE PATIENT DISCONNECTED THE PATIENT LINE FROM THE TRANSFER SET. THE LOT NUMBER IS UNKNOWN; THEREFORE A BATCH REVIEW CANNOT BE PERFORMED. A LABELING REVIEW OF THE HOMECHOICE APD SYSTEMS PATIENT AT-HOME GUIDE ISSUED WAS FOUND TO BE ADEQUATE FOR THE USE ERROR(S) IN THE COMPLAINT. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING A SYSTEM ERROR (SE) 2240 (AIR IN LINE) ALARM THAT APPEARED ON THE HOMECHOICE (HC) DISPLAY DURING DRAIN 1 OF 4. THE HOME PATIENT (HP) STATED THAT SHE DISCONNECTED DURING DWELL 1. THE HP DID NOT PRESS STOP TO PAUSE THERAPY HP THEN RECONNECTED AFTER THE ERROR OCCURRED. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE ALARM. THE HP WOULD NOTIFY THE NURSE. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AT THE TIME OF THE INITIAL REPORT. DURING A FOLLOW UP, THE HOME PATIENT STATED THAT THEY DID NOT DEVELOP ANY ADVERSE REACTIONS OR REQUIRE ANY MEDICAL INTERVENTION. THE HOME PATIENT ALSO STATED THEY ARE CURRENTLY PERFORMING THERAPY WITHOUT ANY COMPLICATIONS. THIS WAS AN ISOLATED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOMATED PD SET W/CASSETTE4 PRONG SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 HC CYCLER