FDA Adverse Event Injury Summary report: N

HAHN TAPERED IMPLANT Ø5.0 X 8 MM

MDR report key: 19705661 · Received July 9, 2024

Report

Report Number
3011649314-2024-00530
Event Type
Injury
Date Received
July 9, 2024
Date of Event
March 18, 2024
Report Date
August 7, 2024
Manufacturer
PRISMATIK DENTALCRAFT, INC.
Product Code
DZE
PMA / PMN Number
K143353
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION HAS BEEN COMPLETED; FOLLOWING IS THE DEVICE ANALYSIS CONCLUSION: DHR RESULTS THE DHR WAS REVIEWED FOR LOT#6200492 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. TS - 07/02/2024. STOCK PRODUCT REVIEWED RESULTS THE STOCK PRODUCT FOR HAHN TAPERED IMPLANT LOT#6125531 IS NOT APPLICABLE FOR REVIEW SINCE THE ISSUE IS CUSTOMER RELATED. INVESTIGATION METHODS/RESULTS THE DEVICE WAS RETURNED BUT NOT IN ORIGINAL PACKAGE. THE IMPLANT WAS VERIFIED TO BE A HAHN TAPERED IMPLANT Ø5 X 8 MM (70-1154-IMP0014) USING RADIOGRAPHIC TEMPLATE (PK-209-062515). THERE WAS NO DEFECT OR NON-CONFORMITY OBSERVED AND THE THREADS WERE INTACT. MATTER WAS OBSERVED IN THE THREADING OF THE IMPLANT. (SEE ATTACHED IMAGES). THE COMPLAINT IS VERIFIED BASED ON THE RETURNED PART(S) BUT CANNOT CONFIRM THE FAILURE MODE. THERE WAS NO EVIDENCE FOUND THAT INDICATED THAT THE REPORTED ISSUE WAS CAUSED BY THE DEVICE ITSELF. TS - 07/08/2024. ROOT CAUSE "FAILURE TO OSSEOINTEGRATE" IS A COMMON COMPLAINT IN REGARD TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES. ALTHOUGH THE ROOT CAUSE FOR FAILURE TO OSSEOINTEGRATE IS INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE THE LOSS OF PRIMARY STABILITY AT THE OSTEOTOMY SITE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT, OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PREMATURE LOADING, PATIENT'S HEALTH, PERI-IMPLANTITIS, SMOKING, AND LACK OF ORAL HYGIENE MAY ALSO BE CONTRIBUTING FACTORS. IFU 570 REV 4 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE CONTRAINDICATIONS SECTION: "HAHN TAPERED IMPLANTS SHOULD NOT BE PLACED IN PATIENTS DISCOVERED TO BE MEDICALLY UNFIT FOR THE INTENDED TREATMENT. PRIOR TO CLINICAL INTERVENTION, PROSPECTIVE PATIENTS MUST BE THOROUGHLY EVALUATED FOR ALL KNOWN RISK FACTORS AND CONDITIONS RELATED TO ORAL SURGICAL PROCEDURES AND SUBSEQUENT HEALING. CONTRAINDICATIONS INCLUDE BUT ARE NOT LIMITED TO: VASCULAR CONDITIONS, UNCONTROLLED DIABETES, CLOTTING DISORDERS, ANTICOAGULANT THERAPY, METABOLIC BONE DISEASE, CHEMOTHERAPY OR RADIATION THERAPY, CHRONIC PERIODONTAL INFLAMMATION, INSUFFICIENT SOFT TISSUE COVERAGE, METABOLIC OR SYSTEMIC DISORDERS ASSOCIATED WITH WOUND AND/OR BONE HEALING, USE OF PHARMACEUTICALS THAT INHIBIT OR ALTER NATURAL BONE REMODELING, ANY DISORDERS WHICH INHIBIT A PATIENT'S ABILITY TO MAINTAIN ADEQUATE DAILY ORAL HYGIENE, UNCONTROLLED PARAFUNCTIONAL HABITS, INSUFFICIENT HEIGHT AND/OR WIDTH OF BONE, AND INSUFFICIENT INTERARCH SPACE." IFU 570 REV 4 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE SURGICAL PROCEDURES UNDER THE PRECAUTION SECTION: "MINIMIZING TISSUE DAMAGE IS CRUCIAL TO SUCCESSFUL IMPLANT OSSEOINTEGRATION. IN PARTICULAR, CARE SHOULD BE TAKEN TO ELIMINATE SOURCES OF INFECTION, CONTAMINANTS, SURGICAL AND THERMAL TRAUMA. RISK OF OSSEOINTEGRATION FAILURE INCREASES AS TISSUE TRAUMA INCREASES. FOR BEST RESULTS, PLEASE OBSERVE THE FOLLOWING PRECAUTIONS: ALL DRILLING PROCEDURES SHOULD BE PERFORMED AT 2000 RPM OR LESS UNDER CONTINUAL, COPIOUS IRRIGATION. ALL SURGICAL INSTRUMENTS USED MUST BE IN GOOD CONDITION AND SHOULD BE USED CAREFULLY TO AVOID DAMAGE TO IMPLANTS OR OTHER COMPONENTS. IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO." IFU 570 REV 4 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE WARNING SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION." IFU 570 REV 4 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE WARNING SECTION: "THE IMPLANT SITE SHOULD BE INSPECTED FOR ADEQUATE BONE BY RADIOGRAPHS, PALPATIONS AND VISUAL EXAMINATION. DETERMINE THE LOCATION OF NERVES AND OTHER VITAL STRUCTURES AND THEIR PROXIMITY TO THE IMPLANT SITE BEFORE ANY DRILLING TO AVOID POTENTIAL INJURY, SUCH AS PERMANENT NUMBNESS TO THE LOWER LIP AND CHIN." 07/08/2024 .THE DEVICE WAS RETURNED AND IS PART OF CAPA 24-005 FOR BACKLOG AND OR LOST DEVICES.

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED FOR ANALYSIS; HOWEVER, THE DEVICE EVALUATION IS PENDING. AT THE COMPLETION OF THE INVESTIGATION A SUPPLEMENTAL REPORT WILL BE SUBMITTED WITH THE ANALYSIS CONCLUSION. COMPLAINT HISTORY AND PRODUCT HISTORY RECORDS WERE REVIEWED; DOCUMENTATION INDICATES THE PRODUCT MET RELEASE CRITERIA. THE PROBABLE ROOT CAUSE OF THE EVENT HAS NOT YET BEEN IDENTIFIED. MANUFACTURER INTERNAL REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

A HEALTHCARE PROFESSIONAL REPORTED A HAHN TAPERED IMPLANT FAILED TO OSSEOINTEGRATE BEFORE THE SECOND STAGE SURGERY OF TOOTH NUMBER 3. IT WAS REPORTED THAT THE HEALTHCARE PROVIDER DID NOT OBSERVE ANY PATIENT SYMPTOMS. THE DEVICE WAS REMOVED, AND IT WAS NOT REPLACED, AND NO ADDITIONAL PROCEDURES WERE PERFORMED. IT WAS REPORTED THAT AT THE TIME OF THE SURGICAL PROCEDURE THE PATIENT'S BONE QUALITY WAS TYPE IV WITH GOOD ORAL HYGIENE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
285736 HAHN TAPERED IMPLANT Ø5.0 X 8 MM HAHN TAPERED IMPLANT SYSTEM DZE PRISMATIK DENTALCRAFT, INC. 70-1154-IMP0014 6200492

Patients

Seq Age Sex Outcome Treatment
1 42 YR Male Required Intervention