FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 1970489 · Received January 19, 2011

Report

Report Number
3002158293-2011-00059
Event Type
Malfunction
Date Received
January 19, 2011
Date of Event
December 26, 2010
Report Date
January 17, 2011
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (SERVICE CODE 204) HAS BEEN CONFIRMED. UPON EVALUATION, THE TRUNK CABLE CONNECTOR WAS BROKEN AND THE BLUE WIRE WAS CUT. THE CAUSE FOR THE BROKEN CONNECTOR AND DAMAGED WIRE CANNOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY THE RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE BROKEN CONNECTOR. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) OLD FEMALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT HER MONITOR IS DISPLAYING SERVICE CODE 204. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 45 YR