FDA Adverse Event Malfunction Summary report: N

INRATIO2

MDR report key: 1970459 · Received January 18, 2011

Report

Report Number
2027969-2011-00123
Event Type
Malfunction
Date Received
January 18, 2011
Date of Event
December 7, 2010
Report Date
January 17, 2011
Manufacturer
ALERE SAN DIEGO, INC.
Product Code
GJS
PMA / PMN Number
K072727
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION PENDING.

Description of Event or Problem · 1

CALLER REPORTS POWER SUPPLY CORD AT THE INDICATOR LIGHT IS SLIGHTLY MELTED AND NO LONGER POWERS ON METER. BATTERIES THAT MAY NOT BE NEW ALSO FAILED TO POWER ON METER. SITE HAS BACK UP GENERATOR THAT HAS BEEN RUNNING A LOT LATELY AND MAY BE CAUSING SURGES IN POWER. REPLACEMENTS POWER CORD SENT, BUT THIS DID NOT RESOLVE THE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INRATIO2 PROTHROMBIN TIME TEST GJS ALERE SAN DIEGO, INC. 200457 NG

Patients

Seq Age Sex Outcome Treatment
1