FDA Adverse Event Malfunction Summary report: N

AFFINITY BED

MDR report key: 1970448 · Received January 18, 2011

Report

Report Number
1824206-2011-00275
Event Type
Malfunction
Date Received
January 18, 2011
Date of Event
December 27, 2010
Report Date
December 27, 2010
Manufacturer
HILL-ROM, INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THAT THE GAS SPRINGS WERE FROZEN AND COULD NOT BE MOVED. HE REPLACED THE GAS SPRINGS AND TRENDELENBURG FUNCTIONED PROPERLY.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE TRENDELENBURG FUNCTION IS NOT OPERATING WHEN THE TREND HANDLES WERE PUSHED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY BED BIRTHING BED HDD HILL-ROM, INC. 3600

Patients

Seq Age Sex Outcome Treatment
1