IMP,TSV,MCOL MG,3.7MM,11.5MML
Report
- Report Number
- 0002023141-2024-02295
- Event Type
- Malfunction
- Date Received
- July 9, 2024
- Date of Event
- June 11, 2024
- Report Date
- November 13, 2024
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024019737
- PMA / PMN Number
- K111889
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ZIMVIE DID NOT RECEIVE ONE (1) TSVMB11, (IMP,TSV,MCOL MG,3.7MM,11.5MML) FOR EVALUATION. VISUAL INSPECTION COULD NOT BE PERFORMED. THE INVESTIGATION HAS BEEN PERFORMED BASED ON THE AVAILABLE INFORMATION USING THE ITEM NUMBER, DEVICE HISTORY RECORD (DHR) REVIEW, AND RISK MANAGEMENT FILE (RMF). FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE & EVENT. THIS COMPLAINT REFERS TO THE SPECIFIC DEVICE BEING INVESTIGATED FOR THIS COMPLAINT RECORD. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1243032. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1243032 FOR SIMILAR EVENTS AND ONE (1) OTHER COMPLAINT WAS IDENTIFIED (B)(4). REVIEW COMPLETED UTILIZING KEYWORDS: ¿FRACTURE IMPLANT¿. THE CUSTOMER DID NOT SUBMIT IMAGES FOR THE REPORTED EVENT. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW AS PER (B)(4), THE MOST LIKELY ROOT CAUSES DETERMINED FROM THE INVESTIGATION ARE LONG-TERM PARAFUNCTIONAL HABITS (E.G., CLENCHING, BRUXISM, AND OVERLOADING) OF THE PATIENT OVER THE IMPLANTATION PERIOD ¿ PATIENT FACTORS. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION COULD NOT BE VERIFIED. WITHOUT DEVICE RECEIPT, THE REPORTED EVENT IS NON-VERIFIABLE. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION. THE FOLLOWING SECTIONS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT, G3: DATE RECEIVED BY MANUFACTURER, G6: CHECKED "FOLLOW-UP", H3: CHANGED "YES" TO "NO", H6: ENTERED EVALUATION CODES, H10: ADDED MANUFACTURER NARRATIVE.
ZIMVIE COMPLAINT NUMBER (B)(4). A2: AGE AND DATE OF BIRTH UNKNOWN / NOT PROVIDED. A3: GENDER UNKNOWN / NOT PROVIDED. A4: PATIENT WEIGHT UNKNOWN / NOT PROVIDED.
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
IT WAS REPORTED THAT THE PATIENT PRESENTED TO DOCTOR WITH LOOSE SCREW AT #3, IT IS DISCOVERED THAT IMPLANT NECK IS FRACTURED. PATIENT LEAVES IN HEALING ABUTMENT AS NEXT STEPS ARE REVIEWED WITH LAB, SURGEON, AND RESTORING CLINICIAN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 258349 | IMP,TSV,MCOL MG,3.7MM,11.5MML | DENTAL IMPLANT | DZE | ZIMMER DENTAL | 1243032 | 00889024019737 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Prefer Not To Disclose |