FDA Adverse Event Malfunction Summary report: N

IMP,TSV,MCOL MG,3.7MM,11.5MML

MDR report key: 19704122 · Received July 9, 2024

Report

Report Number
0002023141-2024-02295
Event Type
Malfunction
Date Received
July 9, 2024
Date of Event
June 11, 2024
Report Date
November 13, 2024
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019737
PMA / PMN Number
K111889
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE DID NOT RECEIVE ONE (1) TSVMB11, (IMP,TSV,MCOL MG,3.7MM,11.5MML) FOR EVALUATION. VISUAL INSPECTION COULD NOT BE PERFORMED. THE INVESTIGATION HAS BEEN PERFORMED BASED ON THE AVAILABLE INFORMATION USING THE ITEM NUMBER, DEVICE HISTORY RECORD (DHR) REVIEW, AND RISK MANAGEMENT FILE (RMF). FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE & EVENT. THIS COMPLAINT REFERS TO THE SPECIFIC DEVICE BEING INVESTIGATED FOR THIS COMPLAINT RECORD. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1243032. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1243032 FOR SIMILAR EVENTS AND ONE (1) OTHER COMPLAINT WAS IDENTIFIED (B)(4). REVIEW COMPLETED UTILIZING KEYWORDS: ¿FRACTURE IMPLANT¿. THE CUSTOMER DID NOT SUBMIT IMAGES FOR THE REPORTED EVENT. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW AS PER (B)(4), THE MOST LIKELY ROOT CAUSES DETERMINED FROM THE INVESTIGATION ARE LONG-TERM PARAFUNCTIONAL HABITS (E.G., CLENCHING, BRUXISM, AND OVERLOADING) OF THE PATIENT OVER THE IMPLANTATION PERIOD ¿ PATIENT FACTORS. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION COULD NOT BE VERIFIED. WITHOUT DEVICE RECEIPT, THE REPORTED EVENT IS NON-VERIFIABLE. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION. THE FOLLOWING SECTIONS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT, G3: DATE RECEIVED BY MANUFACTURER, G6: CHECKED "FOLLOW-UP", H3: CHANGED "YES" TO "NO", H6: ENTERED EVALUATION CODES, H10: ADDED MANUFACTURER NARRATIVE.

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). A2: AGE AND DATE OF BIRTH UNKNOWN / NOT PROVIDED. A3: GENDER UNKNOWN / NOT PROVIDED. A4: PATIENT WEIGHT UNKNOWN / NOT PROVIDED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT PRESENTED TO DOCTOR WITH LOOSE SCREW AT #3, IT IS DISCOVERED THAT IMPLANT NECK IS FRACTURED. PATIENT LEAVES IN HEALING ABUTMENT AS NEXT STEPS ARE REVIEWED WITH LAB, SURGEON, AND RESTORING CLINICIAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
258349 IMP,TSV,MCOL MG,3.7MM,11.5MML DENTAL IMPLANT DZE ZIMMER DENTAL 1243032 00889024019737

Patients

Seq Age Sex Outcome Treatment
1 NA Prefer Not To Disclose