FDA Adverse Event Malfunction Summary report: N

NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE

MDR report key: 1970400 · Received January 18, 2011

Report

Report Number
1811755-2011-00153
Event Type
Malfunction
Date Received
January 18, 2011
Date of Event
December 17, 2010
Report Date
December 20, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
FYD
PMA / PMN Number
K012991
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE TECHNICIAN EVALUATED THE PLUG AND REMOVED THE COVER AND FOUND THAT THE BROWN WIRE WAS MELTED. IT APPEARED THAT THE USER WAS PULLING THE PLUG FROM THE WIRE TO REMOVE IT FROM THE WALL. THERE WAS A LOT OF WIRE SHOWING AT THE PLUG END. THE WIRES WERE REPAIRED AND THE PLUG WAS REPAIRED. THE DEVICE WAS PASSED ALL TESTS AND WAS RETURNED TO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A SERVICE REPAIR VISIT, THE FIELD SERVICE TECH FOUND THAT THE POWER PLUG HAD A MELTED BROWN WIRE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE AIR-HANDLING APPARATUS FOR A SURGICAL OPERATING RO FYD STRYKER INSTRUMENTS KALAMAZOO 0700001000

Patients

Seq Age Sex Outcome Treatment
1 UNK