FDA Adverse Event
Malfunction
Summary report: N
ADJ PIN COLLET 2.0-3.2MM
MDR report key: 1970387
·
Received January 18, 2011
Report
- Report Number
- 1811755-2011-00171
- Event Type
- Malfunction
- Date Received
- January 18, 2011
- Date of Event
- January 6, 2011
- Report Date
- January 6, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- KIJ
- PMA / PMN Number
- K972367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR INVESTIGATION. BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO NOT HOLD THE PIN.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE COLLET WOULD NOT HOLD THE SMALL PIN DURING TESTING IN HOUSE. THERE WAS NO PATIENT INVOLVEMENT AND NO ALLEGATION OF ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.