FDA Adverse Event Malfunction Summary report: N

ADJ PIN COLLET 2.0-3.2MM

MDR report key: 1970387 · Received January 18, 2011

Report

Report Number
1811755-2011-00171
Event Type
Malfunction
Date Received
January 18, 2011
Date of Event
January 6, 2011
Report Date
January 6, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
KIJ
PMA / PMN Number
K972367
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR INVESTIGATION. BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO NOT HOLD THE PIN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE COLLET WOULD NOT HOLD THE SMALL PIN DURING TESTING IN HOUSE. THERE WAS NO PATIENT INVOLVEMENT AND NO ALLEGATION OF ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADJ PIN COLLET 2.0-3.2MM INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR KIJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK