FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 1970362 · Received January 18, 2011

Report

Report Number
1811755-2011-00150
Event Type
Malfunction
Date Received
January 18, 2011
Date of Event
December 29, 2010
Report Date
December 29, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED FOR EVAL AND OVERHEATING WAS DUPLICATED DURING THE QUALITY INVESTIGATION TESTING. FURTHER EVAL REVEALED A BROKEN BALANCES ROTOR AND CORE DRIVESHAFT. THESE PARTS AND OTHER COMPONENT PARTS WERE REPLACED, CLEANED AND LUBED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A STRYKER CORE IMPACTION DRILL OVERHEATED DURING A WISDOM TOOTH EXTRACTION. NO ADVERSE CONSEQUENCES WERE REPORTED AS A RESULT OF THIS EVENT. ANOTHER DRILL WAS USED TO COMPLETE THE PROCEDURE WITHOUT ANY PROCEDURE DELAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE IMPACTION DRILL BONE CUTTING INSTRUMENTS AND ACCESSORIES DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK