FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1970307 · Received January 18, 2011

Report

Report Number
2531779-2011-00325
Event Type
Malfunction
Date Received
January 18, 2011
Report Date
December 19, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED INTACT WITH NO LIFTING OR PEELING OBSERVED, THE DOWN/OK KEYPAD BUTTONS INTERMITTENTLY RESPONDED TO USER INPUTS DURING TESTING, THE UP/CONTRAST KEYPAD BUTTONS REQUIRED EXCESSIVE FORCE TO ACTIVATE DURING TESTING, IT WAS OBSERVED THAT THE CONTRAST KEY CONTACT WAS INVERTED, THE UP/DOWN/OK KEY CONTACTS BECAME INVERTED WHEN PRESSED AND DID NOT ALWAYS SPRING BACK, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE UP ARROW KEYPAD BUTTON INTERMITTENTLY NOT RESPONDING WHEN PRESSED. THE PATIENT CLAIMED THE BUTTON NEEDS TO BE PRESSED SEVERAL TIMES AND WITH ADDITIONAL FORCE FOR A RESPONSE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment