GMK PRIMARY TOTAL KNEE SYSTEM
Report
- Report Number
- 3005180920-2010-00014
- Event Type
- Injury
- Date Received
- January 14, 2011
- Date of Event
- June 14, 2010
- Report Date
- January 11, 2011
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- JWH
- PMA / PMN Number
- K090988
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- NOT APPLICABLE
Narratives
DOCUMENT REVIEW: GMK TIBIAL INSERT FIXED POSTERO STABILIZED - (B)(4)/LOT. 080172. DOCUMENT REVIEW WAS CARRIED OUT ON THE LOT 080172 (15 PIECES PRODUCED): ALL PARAMETERS WERE FOUND TO BE CONFORMING TO SPECIFICATIONS VALID AT THE TIME OF MANUFACTURING. THE WASHING CYCLE FOR METAL COMPONENTS WAS RUN ACCORDING TO SPECIFICATIONS AND NO ALARM OR DEVIATION OCCURRED DURING OPERATION. THE STERILIZATION CYCLE WAS PERFORMED ACCORDING TO SPECIFICATIONS VALID AT THE TIME OF PRODUCTION. ABOUT 6 INSERTS BELONGING TO THIS LOT HAVE BEEN IMPLANTED AND NO INCIDENTS HAVE BEEN REPORTED UP TO NOW. THE REVISION SURGERY IS ASSOCIATED TO AN INFECTION EVENT. ON THE BASIS OF THE BATCH RECORD REVIEW AND THE ABSENCE OF INFECTION EVENTS ON THE OTHER PIECES OF THE SAME LOT, A DEVICE INVOLVEMENT IS HIGH UNLIKELY.
THERE WAS AN EARLY STAGE INFECTION AFTER THE FIRST SURGERY AND THE SURGEON CHANGED THE INSERT IN A REVISION SURGERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GMK PRIMARY TOTAL KNEE SYSTEM | TIBIAL PS FIXED INSERT | JWH | MEDACTA INTERNATIONAL SA | NA | 080172 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Required Intervention |