FLEXTEND II
Report
- Report Number
- 2124215-2011-01265
- Event Type
- Injury
- Date Received
- January 25, 2011
- Date of Event
- November 19, 2010
- Report Date
- November 19, 2010
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. INITIAL INSPECTION NOTED THAT THERE WAS BLOOD INFILTRATION IN THE HELIX HOUSING WHICH INHIBITED THE HELIX FROM EXTENDING OR RETRACTING. THERE WAS ALSO CLAVICULAR INSULATION ABRASION APPROXIMATELY 161 TO 168 MM FROM THE TERMINAL PIN. IN ADDITION, THE CONDUCTOR COILS WERE DEFORMED 18 MM FROM THE TERMINAL PIN AND THERE WAS A HOLE FOUND IN THE INSULATION AT THIS SITE, MOST LIKELY DUE TO THE STYLET. THERE WERE ALSO CUTS IN THE INSULATION WHERE THE SUTURE SLEEVE HAD BEEN LOCATED. NEXT, RESISTANCE TESTING WAS COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE. MEASUREMENTS THROUGHOUT THIS TEST WAS WITHIN NORMAL LIMITS. THE LEAD DID NOT PASS THE PRESSURE TESTING WHICH ASSESSES THE INSULATION INTEGRITY DUE TO THE CUTS AND ABRASIONS IN THE INSULATION. LABORATORY ANALYSIS CONFIRMED THE THAT INSULATION WAS DAMAGED AT THE LOCATION WHERE THE LEAD WOULD HAVE BEEN UNDER THE CLAVICLE WHICH MOST LIKELY CONTRIBUTED TO THE FIELD OBSERVATIONS OF NOISE AND OVERSENSING.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD PRESENTED WITH NOISE FROM MYOPOTENTIALS THAT COULD BE RECREATED WITH PATIENT MOVEMENTS. THE NOISE RESULTED IN OVERSENSING BUT NO INAPPROPRIATE SHOCK WAS DELIVERED. IT IS BELIEVE THAT THE INSULATION WAS DAMAGED. NO ADVERSE PATIENT EFFECTS WERE REPORTED. A LEAD REVISION WAS PERFORMED AND THIS RV LEAD WAS SUCCESSFULLY REPLACED. THE EXPLANTED LEAD WAS THEN RETURNED FOR LABORATORY ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLEXTEND II | IMPLANTABLE LEAD | NVN | CPI - DEL CARIBE | 4097 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |