FDA Adverse Event Malfunction Summary report: N

SELUTE

MDR report key: 1970218 · Received January 25, 2011

Report

Report Number
2124215-2011-01031
Event Type
Malfunction
Date Received
January 25, 2011
Date of Event
January 8, 2007
Report Date
August 3, 2012
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE LEAD WAS RETURNED RETURNED IN TWO SEGMENTS, SEVERED 25.5 CENTIMETERS (CM) FROM THE TERMINAL PIN. PORTIONS OF THE EXTRACTING STYLET REMAINED STUCK IN THE DISTAL SEGMENT OF THE LEAD. THE COILS WERE STRETCHED AND INSULATION WAS TORN APART AT THE DISTAL REGION OF THE LEAD, LIKELY CAUSED BY THE EXPLANT PROCEDURE. AN X-RAY WAS PERFORMED AND NO EVIDENCE OF A FRACTURE WAS FOUND. RESISTANCE TESTING WAS PERFORMED TO ASSESS THE LEAD ELECTRICAL PERFORMANCE. MEASUREMENTS THROUGH THIS TEST WERE WITHIN NORMAL LIMITS. LABORATORY ANALYSIS COULD NOT CONFIRM THE CLINICAL OBSERVATIONS.

Additional Manufacturer Narrative · 1

A NEW LEAD WAS SUCCESSFULLY IMPLANTED. THE ABANDONED LEAD WILL NOT BE RETURNED FOR ANALYSIS THEREFORE BOSTON SCIENTIFIC CANNOT CONFIRM THIS CLINICAL OBSERVATION. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED INDICATING THIS LEAD WAS EXPLANTED DURING A RECENT DEVICE REPLACEMENT PROCEDURE. THE LEAD WAS RETURNED FOR ANALYSIS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING THE ELECTIVE DEVICE REPLACEMENT PROCEDURE, THIS ATRIAL LEAD WAS ABANDONED SURGICALLY DUE TO POOR R-WAVE MEASUREMENTS. NO ADVERSE PATIENT EFFECTS WERE REPORTED. FOUR YEARS LATER, OUR COMPANY RECEIVED INFORMATION FROM THE PATIENT THAT ONE OF HER LEADS HAD FRACTURED. AGAIN, NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SELUTE IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4285

Patients

Seq Age Sex Outcome Treatment
1 57 YR (B)(4)| (B)(4)| (B)(4)| (B)(4)| (B)(4)| (B)(4)