PUMP,COLLEAGUE SINGLE CHANNELCOLOUR
Report
- Report Number
- 6000001-2011-00429
- Event Type
- Malfunction
- Date Received
- January 25, 2011
- Date of Event
- June 4, 2010
- Report Date
- July 28, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Removal / Correction Number
- 6000001-12/1/08-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
THE CONDITION OF BATTERY DEPLETED SET CONFIRMED THROUGH THE EVENT HISTORY. THE CONDITION IS DUE TO DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND HARNESS WERE REPLACED TO CORRECT THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A FOLLOW UP MEDWATCH WILL BE SUBMITTED. (B)(4).
BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4). THE DEVICE WAS PREVIOUSLY SERVICED FOR THE REPORTED CONDITION. (B)(4).
CUSTOMER FROM BAXTER (B)(6) ORIGINALLY REPORTED DAMAGED BATTERY ON A COLLEAGUE INFUSION PUMP. IT IS UNKNOWN WHEN THE REPORTED CONDITION OCCURRED. DURING A REVIEW OF THE PUMP'S EVENT HISTORY BY BAXTER (B)(6) PERSONNEL, THE PUMP WAS FOUND TO HAVE EXPERIENCED A BATTERY DEPLETED SET WHICH INTERRUPTED DELIVERY. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE UTILIZES USER INTERFACE MODULE MASTER SOFTWARE VERSION 6.13.92 CATEGORIZED AS REMEDIATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP,COLLEAGUE SINGLE CHANNELCOLOUR | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |