CLINAC
Report
- Report Number
- 2916710-2010-00160
- Event Type
- Malfunction
- Date Received
- January 14, 2011
- Date of Event
- December 30, 2010
- Report Date
- December 30, 2010
- Manufacturer
- VARIAN MEDICAL SYSTEMS, INC.
- Product Code
- IYE
- PMA / PMN Number
- K904364
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
A VARIAN SERVICE TECHNICIAN PUT THE COVER BACK ONTO THE MACHINE AND VERIFIED IT WAS SECURE FUNCTIONING PROPERLY. VARIAN'S INVESTIGATION REVEALED THAT DURING TREATMENT, THE CENTER THROAT COVER (FIBERGLASS COVER) IS AT ITS LONGEST WHEN THE GANTRY ANGLE IS AT EITHER 90 DEGREES OR 270 DEGREES, SHADOWING OVER THE PT. A LOOSE OR INCORRECTLY APPLIED FASTENER WOULD ENABLE THE LATCHING MECHANISMS TO FAIL, ALLOWING THE THROAT COVER TO FALL ONTO THE PT. THE FIBERGLASS COVER WEIGHS APPROXIMATELY 17 LBS, THE CENTER OF THE COVER IS AT ISO-CENTER (CLOSE TO THE LEVEL OF THE PT) AND THE PART OF THE COVER THAT COULD HIT THE PT IS A FLAT, SMOOTH SURFACE. THE HAZARD ANALYSIS DETERMINED THAT IF THIS SCENARIO OCCURRED, IT COULD CAUSE A SUPERFICIAL CUT AND/OR BRUISE TO THE PT'S HEAD. VARIAN HAS PREVIOUSLY PROVIDED A CUSTOMER TECHNICAL BULLETIN TO ELABORATE ON HOW TO LATCH THE CENTER THROAT COVER PROPERLY. IN ADDITION, A DESIGN IMPROVEMENT TO THE LATCHING MECHANISM WAS CUT-IN TO MANUFACTURING IN AUGUST 2006. THE FIBERGLASS COVER WAS REPLACED AND REINSTALLED AT THIS SITE. FURTHER ACTIONS WILL BE ADDRESSED IN VARIAN'S CAPA PROCESS. NO ADDITIONAL FOLLOW-UP TO THIS MDR IS ANTICIPATED.
WHILE PT WAS BEING TREATED, THE THROAT COVER FELL OFF THE MACHINE AND THE PT WAS STRUCK IN THE HEAD. THE DOCTOR REPORTED THAT THE PT WAS NOT INJURED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CLINAC | LINEAR ACCELERATOR | IYE | VARIAN MEDICAL SYSTEMS, INC. | 2100C/D |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |