FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 1970105 · Received January 14, 2011

Report

Report Number
1720753-2011-00291
Event Type
Malfunction
Date Received
January 14, 2011
Date of Event
December 17, 2010
Report Date
January 14, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE GPOS COMPUTER AND THE HARD DRIVE. RELOADED SOFTWARE, NO ERRORS AT BOOT UP AND DURING SYSTEM OPERATION. SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM DISPLAYED A CORRUPTED FILES ERROR MESSAGE ON BOOT UP AND DELETED IMAGES FROM A PREVIOUS CASE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1