VIVERA RETAINERS
Report
- Report Number
- 2953749-2024-01724
- Event Type
- Injury
- Date Received
- July 9, 2024
- Date of Event
- March 12, 2024
- Report Date
- July 9, 2024
- Manufacturer
- ALIGN TECHNOLOGY, INC.
- Product Code
- DYT
- UDI-DI
- 00816063020028
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
REPORTED EVENT WAS CONFIRMED BY REVIEW OF PHOTOGRAPHS PROVIDED. DEVICE HISTORY RECORD (DHR) WAS REVIEWED FOR THE PART AND LOT NUMBERS, AND NO DISCREPANCIES WERE FOUND. PRODUCTS WERE MANUFACTURED ACCORDING TO INTERNAL AND EXTERNAL REQUIREMENTS. THE CURRENT INSTRUCTIONS FOR USE (IFU) CONTAINS THE FOLLOWING: "GENERAL RISKS TO PATIENTS - GUMS, CHEEKS, OR LIPS MAY BE SCRATCHED OR IRRITATED BY THE RETAINER". NO CONCLUSIVE EVIDENCE HAS BEEN PROVIDED THAT SUPPORTS OR OPPOSES THAT THE VIVERA RETAINERS CAUSED OR CONTRIBUTED TO THE REPORTED SYMPTOM OF SEVERE DYSPLASIA. THIS EVENT IS BEING FILED AS AN MDR PER 21 CFR 803 AS THE TREATING DOCTOR REPORTED THAT THE PATIENT HAD THE SYMPTOM OF SEVERE DYSPLASIA ON TONGUE WHICH REQUIRED SURGERY TO REMOVE AND THE VIVERA RETAINERS WERE BEING USED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ALIGN TECHNOLOGY WILL CONTINUE TO MONITOR FOR TRENDS.
THE TREATING DOCTOR REPORTED THAT THE PATIENT HAD SYMPTOMS OF DISCOMFORT AND A LESION ON THE TONGUE (WHITE SPOT) WHICH GREW LARGER OVER 5 MONTH TIMEFRAME. THE TREATING DOCTOR REPORTED THAT THE PATIENT REQUIRED VISITING THE DENTAL CLINIC TO ALLEVIATE THE REPORTED SYMPTOMS, AND THE PATIENT WAS REFERRED TO STOMATOLOGY FOR BIOPSY AND SURGERY. THE PATIENT REPORTED DISCONTINUING THE USE OF THE VIVERA RETAINERS AND IS CURRENTLY DOING WELL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 250003 | VIVERA RETAINERS | MAINTAINER, SPACE PREFORMED, ORTHODONTIC | DYT | ALIGN TECHNOLOGY, INC. | VIVERA RETAINERS | 158208220 | 00816063020028 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR | Male | Required Intervention| O |