FDA Adverse Event
Malfunction
Summary report: N
PROGRAMMING SOFTWARE
MDR report key: 1970040
·
Received January 14, 2011
Report
- Report Number
- 1644487-2011-00065
- Event Type
- Malfunction
- Date Received
- January 14, 2011
- Date of Event
- December 19, 2010
- Report Date
- December 17, 2010
- Manufacturer
- CYBERONICS INC
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
- Health Professional
- Yes
Narratives
Description of Event or Problem · 1
REPORTER INDICATED, A PATIENT'S SEIZURES WERE INCREASED AND UPON INTERROGATION OF THE VNS, IT WAS NOTED THE OUTPUT CURRENT AND MAGNET OUTPUT CURRENTS WERE SET TO 0MA. THE VNS SETTINGS WERE ADJUSTED BACK TO THE INTENDED SETTINGS. IT WAS SUSPECTED AN INTERRUPTED VNS DIAGNOSTICS TEST OCCURRED AT A PREVIOUS OFFICE VISIT WHICH CAUSED THE VNS SETTINGS TO CHANGE. ATTEMPTS FOR FURTHER INFORMATION ARE IN PROGRESS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROGRAMMING SOFTWARE | LYJ | CYBERONICS INC | MODEL 250 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 23 YR | Female | Required Intervention |