STYLE 15 SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2024-16440
- Event Type
- Injury
- Date Received
- July 9, 2024
- Date of Event
- July 18, 2023
- Report Date
- March 3, 2025
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF RUPTURE WAS RECEIVED ON JULY 18, 2024, WITH LOT NUMBER 1250213. BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS ARE: RUPTURE: OBSERVED, OPENING ASSESSED, AS FOLD FLAW OPENING. AS PER THE INVESTIGATION PROCEDURE: CREASE WEAR ABRASION NICK WAS COMPLETED. AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.
CORRECTION TO G.3. AWARE DATE OF SUPPLEMENTAL MEDWATCH #1. AWARE DATE SHOULD HAVE BEEN LISTED AS 10/AUG/2024.
HEALTHCARE PROFESSIONAL REPORTED, RIGHT SIDE "RUPTURE IMPLANT EXCHANGE". DEVICE WAS EXPLANTED AND REPLACED.
HEALTHCARE PROFESSIONAL REPORTED RIGHT-SIDE "RUPTURE IMPLANT EXCHANGE". DEVICE WAS EXPLANTED AND REPLACED.
HEALTHCARE PROFESSIONAL REPORTED RIGHT-SIDE "RUPTURE IMPLANT EXCHANGE". DEVICE WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 257089 | STYLE 15 SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | 1250213 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 87 YR | Female | Required Intervention |