FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 19698877 · Received July 9, 2024

Report

Report Number
9610877-2024-55165
Event Type
Malfunction
Date Received
July 9, 2024
Date of Event
June 27, 2024
Report Date
July 9, 2024
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
CAL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS NOT DISTRIBUTED IN US SO THAT 510K# IS BLANK. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CFB (COHERENT FIBER BUNDLE) BROKEN. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CFB (COHERENT FIBER BUNDLE). IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE INSERTION FLEXIBLE TUBE BUCKLED, AND THE INSERTION FLEXIBLE TUBE CRUSHED; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT "HR-RPT-0586 (IMAGE FAILURE)" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (BROKEN).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
257021 PENTAX FIBEROPTICINTUBATIONSCOPE 2.4TP RO PLS CAL HOYA CORPORATION PENTAX TOKYO OFFICE FP-7RBS2

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown