FDA Adverse Event Malfunction Summary report: N

THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER

MDR report key: 19696976 · Received July 8, 2024

Report

Report Number
2029046-2024-02292
Event Type
Malfunction
Date Received
July 8, 2024
Date of Event
March 18, 2024
Manufacturer
BIOSENSE WEBSTER INC
Product Code
LPB
UDI-DI
10846835009798
PMA / PMN Number
P030031
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION DETAILS: THE DEVICE WAS RETURNED TO BIOSENSE WEBSTER INC (BWI) FOR EVALUATION AND THE EVALUATION HAS BEEN COMPLETED. A VISUAL INSPECTION AND MAGNETIC SENSOR FUNCTIONALITY EVALUATION OF THE RETURNED DEVICE WAS PERFORMED FOLLOWING BWI PROCEDURES. VISUAL ANALYSIS REVEALED REDDISH MATERIAL AND A HOLE IN THE SURFACE OF THE PEBAX. THE DEVICE WAS CONNECTED TO THE CARTO 3 SYSTEM, IT WAS RECOGNIZED CORRECTLY; HOWEVER, AN ERROR 105 APPEARED ON THE SYSTEM DUE TO AN OPEN CIRCUIT IN THE TIP AREA. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE BATCH NUMBER, AND NO INTERNAL ACTIONS WERE IDENTIFIED. THE MAGNETIC SENSOR ISSUE REPORTED BY THE CUSTOMER WAS CONFIRMED. THE POTENTIAL CAUSE OF THE OPEN CIRCUIT COULD NOT BE DETERMINED. THE POTENTIAL CAUSE OF THE PEBAX DAMAGE COULD BE RELATED TO THE MANIPULATION OF THE DEVICE DURING THE PROCEDURE; HOWEVER, THIS CANNOT BE CONCLUSIVELY DETERMINED. THE CATHETER INSTRUCTIONS FOR USE (IFU) STATE: WHEN CLEANING THE TIP ELECTRODE, BE CAREFUL NOT TO TWIST THE TIP ELECTRODE WITH RESPECT TO THE CATHETER SHAFT; TWISTING MAY DAMAGE THE TIP ELECTRODE BOND AND LOOSEN THE TIP ELECTRODE OR MAY DAMAGE THE CONTACT FORCE SENSOR. IN ADDITION, IN ORDER TO PREVENT DAMAGE TO THE CATHETER TIP, USE THE INSERTION TUBE SUPPLIED WITH THE CATHETER TO ADVANCE OR RETRACT THE CATHETER THROUGH THE HEMOSTASIS VALVE OF THE SHEATH. AFTER INSERTION, SLIDE THE INSERTION TUBE BACK TOWARD THE HANDLE. AS PART OF BIOSENSE WEBSTER¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. EXPLANATION OF CODES: INVESTIGATION FINDINGS: MECHANICAL PROBLEM IDENTIFIED (C07) / INVESTIGATION CONCLUSIONS: UNINTENDED USE ERROR CAUSED OR CONTRIBUTED TO EVENT (D1102) / COMPONENT CODE: SLEEVE (G04115) WERE SELECTED AS RELATED TO THE HOLE IN THE PEBAX IDENTIFIED BY BWI PAL. INVESTIGATION FINDINGS: OPEN CIRCUIT (C0205) / INVESTIGATION CONCLUSIONS: CAUSE NOT ESTABLISHED (D15) / COMPONENT CODE: SENSOR (G03012) WERE SELECTED AS RELATED TO THE CUSTOMER'S REPORTED MAGNETIC SENSOR ISSUE AND OPEN CIRCUIT. IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. MANUFACTURER'S REF. # (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT AN ISCHEMIC VENTRICULAR TACHYCARDIA (ISVT) ABLATION PROCEDURE WITH A THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER FOR WHICH BIOSENSE WEBSTER¿S PRODUCT ANALYSIS LAB (PAL) IDENTIFIED A HOLE ON THE PEBAX. IT WAS INITIALLY REPORTED BY THE CUSTOMER THAT THERE WAS A CATHETER SENSOR ERROR OF ¿MAGNETIC SENSOR APPEARS DISCONNECTED¿. THEY TRIED REPLACING THE CONNECTION CABLE, BUT THE ERROR DID NOT SOLVE. NO FORCE VALUES WERE RECORDED. AFTER REPLACING THE CATHETER, THE ERROR SOLVED AND THE PROCEDURE WAS COMPLETED. THERE WERE NO PATIENT CONSEQUENCES. THE CUSTOMER'S REPORTED MAGNETIC SENSOR ERROR IS NOT CONSIDERED TO BE MDR REPORTABLE SINCE THE POTENTIAL RISK THAT IT COULD CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR DEATH TO THE OPERATOR OR PATIENT IS REMOTE. ON 14-JUN-2024, THE BWI PAL REVEALED THAT A VISUAL INSPECTION OF THE RETURNED DEVICE FOUND A REDDISH MATERIAL AND A HOLE IN THE SURFACE OF THE PEBAX. THESE FINDINGS WERE REVIEWED AND ASSESSED THE ISSUE OF A ¿HOLE¿ IN THE PEBAX AS AN MDR REPORTABLE MALFUNCTION SINCE THE INTEGRITY OF THE DEVICE HAS BEEN COMPROMISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
170048 THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER CARDIAC ABLATION PERCUTANEOUS CATHETER LPB BIOSENSE WEBSTER INC 31128594L 10846835009798

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown