FDA Adverse Event Malfunction Summary report: N

BREMER ADJUSTIBLE RING

MDR report key: 196948 · Received November 12, 1998

Report

Report Number
1035798-1998-00001
Event Type
Malfunction
Date Received
November 12, 1998
Date of Event
August 26, 1998
Report Date
August 27, 1998
Manufacturer
BREMER MEDICAL, INC
Product Code
HAX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE RING FROM THE HALO WAS PUT ON THE PT IN ORDER TO PUT THE PT IN TRACTION PRIOR TO PUTTING ON THE HALO VEST. THE ORTHOTIST HAD DIFFICULTY WHEN HE ATTEMPTED TO PUT ON THE TRACTION BALES BECAUSE THE BOLTS WOULD NOT GAIN A PURCHASE ON THE RING. HE USED A COUPLE OF LONGER BOLTS THAT HE HAD IN HIS POSSESSION. THE TRACTION BALE AND RING FUNCTIONED WELL FOR FIVE DAYS, AFTER WHICH TIME THE TRACTION BALE WAS REMOVED AND THE PT PUT IN AN AIRFLO VEST. ONE HALF DAY LATER THE PT NOTED THAT HE COULD MOVE HIS HEAD. WHEN THE ORTHOTIST TRIED TO TIGHTEN THE BOLTS, THEY STRIPPED THE THREADS IN THE RING. WHEN HE TRIED TO TIGHTEN THE OTHER FOUR BOLTS THAT HOLD THE RING TOGETHER, THEY ALSO STRIPEPD. THE HALO VEST WAS REMOVED AND THE PT WAS PUT INTO ANOTHER SYSTEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BREMER ADJUSTIBLE RING Implant ORTHOTIC BRACE HAX BREMER MEDICAL, INC NA 28761

Patients

Seq Age Sex Outcome Treatment
1 61 YR