FDA Adverse Event Malfunction Summary report: N

FLUSHING PUMP OFP-2 (JP)

MDR report key: 19691400 · Received July 8, 2024

Report

Report Number
9611174-2024-01299
Event Type
Malfunction
Date Received
July 8, 2024
Report Date
September 23, 2024
Manufacturer
KEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD.
Product Code
FEQ
PMA / PMN Number
K100899
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

E2/E3: NO INFORMATION AVAILABLE FOR THE OCCUPATION OF THE INITIAL REPORTER OR WHETHER THEY ARE A HEALTHCARE PROFESSIONAL. THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND THE REPORTED FAILURE WAS CONFIRMED. IN ADDITION, THE FOLLOWING REPORTABLE MALFUNCTIONS WERE IDENTIFIED DURING THE DEVICE EVALUATION: FLUSHING PUMP NOT WORKING. A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: THE EVENT MAY BE CAUSED BY A COMPONENT FAILURE OF ADAPTER PLATE AND CONTROL PANEL. THE CAUSE OF THE ADAPTER PLATE AND CONTROL PANEL FAILURE COULD NOT BE DETERMINED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE STAND-BY SWITCH DID NOT WORK ON THE FLUSHING PUMP. THERE WERE NO REPORTS OF PATIENT HARM.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE STAND-BY SWITCH DID NOT WORK ON THE FLUSHING PUMP. THERE WERE NO REPORTS OF PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
911216 FLUSHING PUMP OFP-2 (JP) OLYMPUS FLUSHING PUMP FEQ KEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD. K10001142

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown