FDA Adverse Event Malfunction Summary report: N

ARCHITECT FREE T4 REAGENT KIT

MDR report key: 19679626 · Received July 5, 2024

Report

Report Number
3005094123-2024-00366
Event Type
Malfunction
Date Received
July 5, 2024
Date of Event
June 15, 2024
Report Date
September 16, 2024
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
CEC
UDI-DI
00380740173678
PMA / PMN Number
K173122
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Additional Manufacturer Narrative · 0

SECTION D4: PRIMARY UDI NUMBER: THIS SECTION WAS CORRECTED FROM (B)(4) TO (B)(4). THE COMPLAINT INVESTIGATION FOR FALSELY ELEVATED ARCHITECT FREE T4 RESULT INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, TICKET TRENDING REVIEW, DEVICE HISTORY RECORD REVIEW, AND LABELING REVIEW. IN ADDITION, IN-HOUSE TESTING OF RETAINED REAGENT KIT WAS ALSO COMPLETED. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. THE DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE. A REVIEW OF THE COMPLAINT TRENDING REPORT DID IDENTIFY AN INCREASE IN COMPLAINTS FOR LOT 61272UD02, HOWEVER, IN-HOUSE PERFORMANCE TESTING WAS COMPLETED WHICH INDICATES THE PRODUCT IS PERFORMING AS EXPECTED. A REVIEW OF THE DEVICE HISTORY RECORD DID NOT IDENTIFY ANY NON-CONFORMANCE'S OR DEVIATIONS WITH LOT NUMBER 61272UD02 AND THE COMPLAINT ISSUE. A REVIEW OF LABELING WAS PERFORMED AND FOUND TO SUFFICIENTLY ADDRESS THE CUSTOMER'S ISSUE. BASED ON THIS INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY WAS IDENTIFIED WITH THE ARCHITECT FREE T4 REAGENT, LOT NUMBER 61272UD02.

Description of Event or Problem · 0

THE CUSTOMER REPORTED FALSELY ELEVATED ARCHITECT FREE T4 RESULT GENERATED ON THE ARCHITECT I2000SR PROCESSING MODULE FOR ONE PATIENT. THE SAMPLE WAS REPEATED AND A NORMAL RESULT WAS OBTAINED. THE FOLLOWING DATA WAS PROVIDED: CUSTOMER¿S REFERENCE RANGE: 9.01 TO 19.05 PMOL/L INITIAL ARCHITECT FREE T4 RESULT = >64 PMOL/L, REPEAT RESULT = 18.14 PMOL/L THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED FALSELY ELEVATED ARCHITECT FREE T4 RESULT GENERATED ON THE ARCHITECT I2000SR PROCESSING MODULE FOR ONE PATIENT. THE SAMPLE WAS REPEATED AND A NORMAL RESULT WAS OBTAINED. THE FOLLOWING DATA WAS PROVIDED: CUSTOMER¿S REFERENCE RANGE: 9.01 TO 19.05 PMOL/L. INITIAL ARCHITECT FREE T4 RESULT = >64 PMOL/L, REPEAT RESULT = 18.14 PMOL/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1133887 ARCHITECT FREE T4 REAGENT KIT RADIOIMMUNOASSAY, FREE THYROXINE CEC ABBOTT IRELAND DIAGNOSTICS DIVISION 61272UD02 00380740173678

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown ARC I2000SR INST, 03M74-02, ISR62378.| ARC I2000SR INST, 03M74-02, ISR62378.