ARCHITECT FREE T4 REAGENT KIT
Report
- Report Number
- 3005094123-2024-00366
- Event Type
- Malfunction
- Date Received
- July 5, 2024
- Date of Event
- June 15, 2024
- Report Date
- September 16, 2024
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- CEC
- UDI-DI
- 00380740173678
- PMA / PMN Number
- K173122
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
SECTION D4: PRIMARY UDI NUMBER: THIS SECTION WAS CORRECTED FROM (B)(4) TO (B)(4). THE COMPLAINT INVESTIGATION FOR FALSELY ELEVATED ARCHITECT FREE T4 RESULT INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, TICKET TRENDING REVIEW, DEVICE HISTORY RECORD REVIEW, AND LABELING REVIEW. IN ADDITION, IN-HOUSE TESTING OF RETAINED REAGENT KIT WAS ALSO COMPLETED. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. THE DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE. A REVIEW OF THE COMPLAINT TRENDING REPORT DID IDENTIFY AN INCREASE IN COMPLAINTS FOR LOT 61272UD02, HOWEVER, IN-HOUSE PERFORMANCE TESTING WAS COMPLETED WHICH INDICATES THE PRODUCT IS PERFORMING AS EXPECTED. A REVIEW OF THE DEVICE HISTORY RECORD DID NOT IDENTIFY ANY NON-CONFORMANCE'S OR DEVIATIONS WITH LOT NUMBER 61272UD02 AND THE COMPLAINT ISSUE. A REVIEW OF LABELING WAS PERFORMED AND FOUND TO SUFFICIENTLY ADDRESS THE CUSTOMER'S ISSUE. BASED ON THIS INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY WAS IDENTIFIED WITH THE ARCHITECT FREE T4 REAGENT, LOT NUMBER 61272UD02.
THE CUSTOMER REPORTED FALSELY ELEVATED ARCHITECT FREE T4 RESULT GENERATED ON THE ARCHITECT I2000SR PROCESSING MODULE FOR ONE PATIENT. THE SAMPLE WAS REPEATED AND A NORMAL RESULT WAS OBTAINED. THE FOLLOWING DATA WAS PROVIDED: CUSTOMER¿S REFERENCE RANGE: 9.01 TO 19.05 PMOL/L INITIAL ARCHITECT FREE T4 RESULT = >64 PMOL/L, REPEAT RESULT = 18.14 PMOL/L THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
THE CUSTOMER REPORTED FALSELY ELEVATED ARCHITECT FREE T4 RESULT GENERATED ON THE ARCHITECT I2000SR PROCESSING MODULE FOR ONE PATIENT. THE SAMPLE WAS REPEATED AND A NORMAL RESULT WAS OBTAINED. THE FOLLOWING DATA WAS PROVIDED: CUSTOMER¿S REFERENCE RANGE: 9.01 TO 19.05 PMOL/L. INITIAL ARCHITECT FREE T4 RESULT = >64 PMOL/L, REPEAT RESULT = 18.14 PMOL/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1133887 | ARCHITECT FREE T4 REAGENT KIT | RADIOIMMUNOASSAY, FREE THYROXINE | CEC | ABBOTT IRELAND DIAGNOSTICS DIVISION | 61272UD02 | 00380740173678 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | ARC I2000SR INST, 03M74-02, ISR62378.| ARC I2000SR INST, 03M74-02, ISR62378. |