FDA Adverse Event Malfunction Summary report: N

AUTOSOFT XC

MDR report key: 19679137 · Received July 5, 2024

Report

Report Number
3003442380-2024-12392
Event Type
Malfunction
Date Received
July 5, 2024
Date of Event
May 16, 2024
Report Date
December 27, 2024
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244016620
PMA / PMN Number
K032854
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INITIAL AND FINAL MDR 1904787 DEVICE 2 OF 4.

Additional Manufacturer Narrative · 0

SUPPLEMENTAL REPORT 01 - MDR (B)(4) - MDR 3003442380-2024-12392. H11: INVESTIGATION SUMMARY: THE COMPLAINT PR.(B)(4) HAS BEEN EVALUATED. THE BATCH 6004389 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. INVESTIGATION PROCESS OF THE COMPLAINT WAS CARRIED OUT IN ACCORDANCE WITH WORK INSTRUCTION GUIDANCE FOR VISUAL TESTING FOR COMPLAINTS AREA VERSION 1 AND WORK INSTRUCTION GUIDANCE FOR FUNCTIONAL TESTING FOR COMPLAINTS AREA VERSION 1 FOR THE CODE "SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE." COMPLAINT INVESTIGATIONS. PHOTO/SAMPLE WAS NOT PROVIDED. IN ORDER TO TEST THE PRODUCT, THE REFERENCE SAMPLES FROM THE LOT HAVE BEEN REQUESTED. THE FOLLOWING TEST WAS PERFORMED: VISUAL TEST ACCORDING TO WORK INSTRUCTION VERSION 1 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL TEST 1 ACCORDING TO WORK INSTRUCTION VERSION 1 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. A BATCH RECORD REVIEW WAS CONDUCTED RESULTING IN THE FOLLOWING: THE LOT 6004389 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION VERSION 108 MANUFACTURED IN THE LINEA 3, ON 28/NOV/2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESS HAD BEEN FULFILLED AND MET THE REQUIREMENTS. A QUERY WAS RUN IN TRACKWISE AGAINST MALFUNCTION CODE "SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE." AND LOT 6004836 AND 8 COMPLAINTS HAVE BEEN REGISTERED IN TW SINCE THE LAST 12 MONTHS. CONCLUSION SUMMARY OF THE RELATED EVENT: AS A RESULT OF THE FOLLOWING: HARM NO REPORTABLE, NO DEFECT ON TESTS, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, 8 COMPLAINTS RECEIVED ON THE LOT IN QUESTION AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKETING SURVEILLANCE (PMS) PRODUCT TRENDS AND MALFUNCTION ACCORDING TO THE ON MARKET QUALITY REVIEW (MQR) PROCEDURE

Description of Event or Problem · 0

UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. ON (B)(6) 2024 , IT WAS REPORTED THAT THE PATIENT FACED INFUSION SET CANNULA WAS KINKED WITHIN 3 HOURS OF INSERTION AT ABDOMEN, WHICH CAUSED THE PATIENT BLOOD GLUCOSE ELEVATED TO HIGH, THEREFORE PATIENT HAD RECEIVED CORRECTION BOLUS VIA PUMP AND CHANGED THE INFUSION SET. THE PATIENT REPLACED INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. UNOMEDICAL DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION AVAILABLE

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
977892 AUTOSOFT XC UNO INSET I 60/6 GREY TCAP 10PK INT FPA UNOMEDICAL A/S 1001680 6004389 05705244016620

Patients

Seq Age Sex Outcome Treatment
1 14 YR Male