FDA Adverse Event Injury Summary report: N

40MM I.D. SIZE LL NEUTRAL LINER USE WITH 58MM O.D. SIZE LL SHELL

MDR report key: 19671654 · Received July 3, 2024

Report

Report Number
0001822565-2024-02235
Event Type
Injury
Date Received
July 3, 2024
Date of Event
June 4, 2024
Report Date
October 17, 2024
Manufacturer
ZIMMER BIOMET, INC.
Product Code
LPH
UDI-DI
00889024149243
PMA / PMN Number
K151448
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL AND/OR CORRECTED INFORMATION. UPDATED: D6A, G3, G6, H2.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL AND/OR CORRECTED INFORMATION. UPDATED: D4, G3, G6, H2, H3, H6. NO PRODUCT WAS RETURNED, OR PICTURES PROVIDED; VISUAL AND DIMENSIONAL EVALUATIONS COULD NOT BE PERFORMED. REVIEW OF THE DEVICE HISTORY RECORDS IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING. REVIEW OF COMPLAINT HISTORY FOUND NO ADDITIONAL RELATED ISSUES FOR THIS ITEM AND THE REPORTED PART AND LOT COMBINATION. MEDICAL RECORDS WERE NOT PROVIDED. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. THIS COMPLAINT CANNOT BE CONFIRMED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 0

(B)(4). D10: 00-8775-040-03 BIOLOX DELTA FEM HEAD, 40MM, +3.5MM 3103810 CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT UNDERWENT A HIP ARTHROPLASTY. SUBSEQUENTLY, THE PATIENT IS UNDERGOING A REVISION SURGERY TO CHANGE THE LINER TO A CONSTRAINED LINER. PATIENT SUFFERS FROM A PREVIOUS PELVIC FRACTURE AND DISLOCATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1283014 40MM I.D. SIZE LL NEUTRAL LINER USE WITH 58MM O.D. SIZE LL SHELL PROSTHESIS, HIPS LPH ZIMMER BIOMET, INC. N/A 65346726 00889024149243

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Hospitalization| R SEE H11 NARRATIVE.