FDA Adverse Event Malfunction Summary report: N

FAST-FIX 360 CURVED NDL DELIVERY SYS

MDR report key: 19668405 · Received July 3, 2024

Report

Report Number
1219602-2024-01432
Event Type
Malfunction
Date Received
July 3, 2024
Date of Event
June 13, 2024
Report Date
July 17, 2024
Manufacturer
SMITH & NEPHEW, INC.
Product Code
GAT
UDI-DI
00885554023077
PMA / PMN Number
K121861
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INTERNAL COMPLAINT REFERENCE (B)(4). H10: THIS COMPLAINT WAS OPENED BY SMITH+NEPHEW TO DOCUMENT A PRODUCT PROBLEM ASSOCIATED WITH A SMITH+NEPHEW DEVICE. THE REPORTED PROBLEM RELATES TO KNOWN INHERENT DEVICE AND/OR PROCEDURAL RISKS THAT ARE APPROPRIATELY DOCUMENTED IN OUR RISK FILES. SMITH+NEPHEW WILL CONTINUE TO MONITOR TRENDS IN ACCORDANCE WITH OUR POST-MARKET SURVEILLANCE PROCESS AND TAKE NECESSARY ACTION AS REQUIRED IF ANTICIPATED SEVERITY AND/OR OCCURRENCE RATES ARE EXCEEDED. SMITH+NEPHEW HAS NO REASON TO SUSPECT THAT THE PRODUCT FAILED TO MEET ANY SPECIFICATIONS AT THE TIME OF MANUFACTURE. BASED ON OUR REVIEW OF ALL CURRENTLY AVAILABLE INFORMATION, WE ARE UNABLE TO IDENTIFY A DEFINITIVE ROOT CAUSE. HOWEVER, AS THE USE OF OUR PRODUCT CANNOT BE EXCLUDED AS A POTENTIAL CAUSE OR CONTRIBUTORY FACTOR TO THE REPORTED ISSUE, WE ARE CONSERVATIVELY SUBMITTING THIS REPORT IN ACCORDANCE WITH APPLICABLE REGULATIONS. IF ADDITIONAL INFORMATION BECOMES AVAILABLE THAT ALTERS THE CONCLUSIONS OF THIS REPORT, A FOLLOW-UP REPORT WILL BE SUBMITTED AS REQUIRED.

Additional Manufacturer Narrative · 0

H3, H6: PART OF THE REPORTED DEVICE WAS RECEIVED FOR EVALUATION. A VISUAL INSPECTION REVEALED THAT IT IS NOT IN ITS ORIGINAL PACKAGING. THE INSERTION DEVICE WAS RETURNED IN THE PRET2/POSTT1 SETTING WITH NO SUTURE OR IMPLANTS RETURNED. THERE IS BIOLOGICAL DEBRIS ON THE RETURNED DEVICE. A FUNCTIONAL EVALUATION WAS PERFORMED ON THE RETURNED DEVICE AND FOUND IT CYCLES AS DESIGNED. A REVIEW OF DEVICE RECORDS SHOWED THERE WERE NO INDICATIONS TO SUGGEST THAT THE PRODUCT DID NOT MEET MANUFACTURING SPECIFICATION UPON RELEASE FOR DISTRIBUTION. A COMPLAINT HISTORY REVIEW FOUND SIMILAR REPORTED EVENTS. A RISK MANAGEMENT REVIEW FOUND THAT THE REPORTED FAILURE AND/OR HARM WAS DOCUMENTED APPROPRIATELY, AND THERE WERE NO INDICATIONS TO SUGGEST THE ANTICIPATED RISK IS NOT ADEQUATE. A CLINICAL EVALUATION STATES THAT AN IMAGE OF THE DEVICE WAS PROVIDED FOR REVIEW; HOWEVER, IT DOES NOT INDICATE A ROOT CAUSE FOR THE REPORT EVENTS. THE ROOT CAUSE COULD NOT BE DETERMINED SINCE THE REPORTED MALFUNCTION COULD NOT BE DUPLICATED DURING THE INVESTIGATION. FACTORS THAT CAN CONTRIBUTE TO THE REPORTED EVENT INCLUDE AN UNINTENDED ACTUATION OF THE DEVICE. BASED ON THIS INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING MENISCUS REPAIR SURGERY, THE FAST-FIX 360 REACHES THE DESIRED SITE ALONG THE 7210977 SLOTTED CANNULA DEVICE. AFTER DEPLOYING T1, WHEN WITHDRAWING THE NEEDLE TO THE WHITE AREA TO PREPARE FOR DEPLOYING T2, IT WAS FOUND THAT T2 HAD FALL OFF AND WAS NO LONGER ON THE INSERTER. THE IMPLANT WAS REMOVED USING HEMOSTATIC FORCEPS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DELAY OF LESS THAN 30 MINUTES USING A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
447626 FAST-FIX 360 CURVED NDL DELIVERY SYS SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE GAT SMITH & NEPHEW, INC. 2110739 00885554023077

Patients

Seq Age Sex Outcome Treatment
1 18 YR Male