PANDA IRES WARMER
Report
- Report Number
- 2112667-2024-03538
- Event Type
- Injury
- Date Received
- July 3, 2024
- Date of Event
- November 20, 2018
- Report Date
- July 3, 2024
- Manufacturer
- DATEX-OHMEDA, INC.
- Product Code
- FMT
- PMA / PMN Number
- K201628
- Removal / Correction Number
- Z-1846-2019 & Z-1847-201
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- RISK MANAGER
- Health Professional
- N
Narratives
GE HEALTHCARE (GEHC) REPORTED A FIELD MODIFICATION FOR THIS ISSUE PER 21 CFR 806 ON 20 MARCH 2019. THE FDA RECALL NUMBER IS Z-1846-2019 & Z-1847-2019. - CUSTOMERS WERE SENT A LETTER EXPLAINING THE ISSUE AND REQUESTING THE CUSTOMER TO INSPECT THE WARMER BEDSIDE PANEL LATCH AREAS. REPLACEMENT OF BROKEN BEDSIDE PANELS WAS PROVIDED BY GEHC. - A SET OF WARNING LABELS WAS SUPPLIED FOR APPLICATION TO THE BEDSIDE PANELS. THESE LABELS WARN THE USER TO NOT USE THE BEDSIDE PANELS FOR MANEUVERING THE WARMER AND INDICATE THE CORRECT METHOD OF MANEUVERING THE WARMER. - AN ADDENDUM TO THE OPERATION AND MAINTENANCE MANUAL WAS PROVIDED EMPHASIZING THE NEED TO CHECK AND ENSURE THAT THE BEDSIDE PANELS AND LATCHES ARE NOT CRACKED, BROKEN, OR DAMAGED BEFORE EVERY PATIENT USE. THE ADDENDUM CONTAINED INSTRUCTIONS TO INCREASE DETECTABILITY OF BROKEN OR CRACKED BEDSIDE PANELS. - ON JUNE 14, 2019 A LETTER WAS SENT TO CUSTOMERS PROVIDING AN ADDITIONAL ADDENDUM, LABELS, AND WALL POSTER. D4 SERIAL NUMBER: NOT PROVIDED.. SECTION E: INITIAL REPORTER / CONTACT INFORMATION NOT PROVIDED. G2 REPORT SOURCE OTHER: MAUDE DATA BASE REPORT # (B)(4). H4 DATE OF DEVICE MANUFACTURE: UNKNOWN AS THE SERIAL NUMBER WAS NOT PROVIDED. LEGAL MANUFACTURER: HCS RESEARCH PARK - 9900 INNOVATION DRIVE USA WAUWATOSA, WI 53226.
PER MAUDE DATA BASE REPORT # (B)(4): "PLASTIC HINGE AT FOOT OF PANDA WARMER CRACKED AND RESULTED IN A FAILURE OF THE END PLASTIC PANELS. THE END PANEL OPENED DUE TO THE CRACKED PLASTIC AND INFANT FELL TO FLOOR."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 959353 | PANDA IRES WARMER | WARMER, INFANT RADIANT | FMT | DATEX-OHMEDA, INC. | M1112198 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 1 DA | Unknown | Hospitalization |