AUTOSOFT XC

MDR report key: 19664033 · Received July 3, 2024

Report

Report Number
3003442380-2024-11392
Event Type
Malfunction
Date Received
July 3, 2024
Date of Event
May 1, 2024
Report Date
July 1, 2024
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244016620
PMA / PMN Number
K032854
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INITIAL AND FINAL MDR 1900950 - MDR 3003442380-2024-11392 - DEVICE 2 OF 4.

Description of Event or Problem · 0

UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. ON (B)(6) 2024, THE PATIENT FACED THAT THE INFUSION SET CANNULA WAS KINKED WITHIN 3 HOURS OF INSERTION AT ABDOMEN, WHICH CAUSE THE PATIENT'S BLOOD GLUCOSE ELEVATED FROM 271 TO 300 MG/DL. THE PATIENT REPLACED INFUSION SET AND RESUMED INSULIN SUCCESSFULLY. UNOMEDICAL DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
602364 AUTOSOFT XC UNO INSET I 60/6 GREY TCAP 10PK INT FPA UNOMEDICAL A/S 1001680 6003525 05705244016620

Patients

Seq Age Sex Outcome Treatment
1 66 YR Female