FDA Adverse Event Malfunction Summary report: N

LIVONGO BLOOD GLUCOSE METER

MDR report key: 19660656 · Received July 2, 2024

Report

Report Number
3011196194-2024-00032
Event Type
Malfunction
Date Received
July 2, 2024
Date of Event
April 23, 2024
Report Date
July 2, 2024
Manufacturer
TELADOC HEALTH, INC.
Product Code
NBW
PMA / PMN Number
K133584
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE PATIENT WAS SENT NEW SUPPLIES TO PERFORM CONTROL SOLUTION TESTS AGAIN. MULTIPLE ATTEMPTS WERE MADE TO REACH THE PATIENT AND WERE UNSUCCESSFUL. WE WERE UNABLE TO RUN CONTROL SOLUTION TESTS AGAIN WITH THE PATIENT, AND THE DEVICE WAS NOT RETURNED FOR THE INVESTIGATION.

Description of Event or Problem · 0

THE PATIENT PERFORMED CONTROL SOLUTION TESTS WITH THE LIVONGO BLOOD GLUCOSE METER AND THE RESULTS CAME BACK OUT OF RANGE TWICE. THE RESULT FOR CONTROL SOLUTION 1 CAME BACK AS 158, AND 216. THE PRE-CALIBRATED RANGE FOR CONTROL 1 WAS 98-147. THE PATIENT DIDN'T RECEIVE MEDICAL ATTENTION AND/OR TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
755827 LIVONGO BLOOD GLUCOSE METER BLOOD GLUCOSE METER NBW TELADOC HEALTH, INC. BG300C

Patients

Seq Age Sex Outcome Treatment
1 45 YR Unknown