INTELLIS
Report
- Report Number
- 3004209178-2024-14040
- Event Type
- Malfunction
- Date Received
- July 2, 2024
- Date of Event
- June 26, 2024
- Report Date
- July 2, 2024
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- LGW
- UDI-DI
- 00763000315467
- PMA / PMN Number
- P840001
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- 003
Narratives
MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
INFORMATION WAS RECEIVED FROM A PATIENT VIA MANUFACTURER REPRESENTATIVE WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS). IT WAS REPORTED THAT THERE WERE HIGH IMPEDANCES: ELECTRODE 2- 31900, 3- 32520, 4-20080. THE REP WAS ABLE TO ADJUST PROGRAMMING AND ELECTRODE CONFIGURATION TO AVOID HIGH IMPEDANCE ELECTRODES. NO SYMPTOMS OR STIMULATION ISSUES REPORTED. ISSUE RESOLVED, BUT THE CAUSE IS UNKNOWN. THE REP REPORTED THEY WOULD SUBMIT ANY NEW INFORMATION THAT BECOMES AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 760040 | INTELLIS | STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR | LGW | MEDTRONIC PUERTO RICO OPERATIONS CO. | 97715 | 00763000315467 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Female |