FDA Adverse Event Malfunction Summary report: N

NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE

MDR report key: 1965786 · Received January 14, 2011

Report

Report Number
1811755-2011-00133
Event Type
Malfunction
Date Received
January 14, 2011
Date of Event
December 16, 2010
Report Date
December 17, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
FYD
PMA / PMN Number
K012991
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE TECH EVALUATED THE ROVER AND FOUND THAT POWER PLUG WAS PARTLY MELTED DUE TO ELECTRICAL CONNECTIONS BECOMING LOOSE. THE POWER PLUG WAS REPLACED AND THE ACCOUNT WAS TRAINED ON NOT PULLING THE PLUG OUT OF THE WALL BY THE CORD.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PLUG ON THE ROVER WAS PARTIALLY MELTED. THERE WAS NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE AIR-HANDLING APPARATUS FOR A SURGICAL OPERATING RO FYD STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK