FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1965185 · Received December 23, 2010

Report

Report Number
2531779-2010-02911
Event Type
Malfunction
Date Received
December 23, 2010
Report Date
November 17, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE 'CONTRAST/UP/OK' BUTTONS ARE RESPONSIVE AND 'DOWN' BUTTON INTERMITTENTLY ARE RESPONSIVE TO USER INPUT. ALL BUTTONS SPRING BACK WHEN PRESSED. AFTER THE KEYPAD WAS REMOVED, AN INTERNAL INSPECTION REVIEWED CONTAMINATION UNDER 'DOWN/CONTRAST' BUTTON CONTACTS. THERE WAS NO CONTAMINATION OBSERVED UNDER 'OK/UP/BUTTON CONTACTS.

Description of Event or Problem · 1

THIS COMPLAINT IS BEING REPORTED DUE TO AN ALLEGED KEYPAD MALFUNCTION. THERE WAS NO REPORT OF PRODUCT MISUSE. THE KEYPAD DID NOT APPEAR TO BE PEELED OR TORN. FURTHERMORE, THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020