VIDEO SYSTEM CENTER
Report
- Report Number
- 3002808148-2024-05975
- Event Type
- Malfunction
- Date Received
- July 1, 2024
- Date of Event
- May 24, 2024
- Report Date
- July 19, 2024
- Manufacturer
- SHIRAKAWA OLYMPUS CO., LTD.
- Product Code
- FDF
- UDI-DI
- 04953170356384
- PMA / PMN Number
- K111756
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. IT HAS BEEN OVER 4 YEARS SINCE THE SUBJECT DEVICE WAS MANUFACTURED. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION, AND THE REPORTABLE MALFUNCTION WAS CONFIRMED. BASED ON THE RESULTS OF THE INVESTIGATION, A DEFINITIVE ROOT CAUSE FOR THE EVENT COULD NOT BE DETERMINED. HOWEVER, IT IS LIKELY THAT THE PHENOMENON ¿FRONT PANEL SWITCHES STOPPED OPERATION¿ OCCURRED DUE TO FAILURE OF THE RECEPTACLE UNIT. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
F24276771-1 : [EVALUATION FOR MDR SUBMISSION BY COE]. EVALUATION DATE: 2024/6/27. EVALUATOR: (B)(6). PURPOSE: PERIODIC INSPECTION. (INTAKE AND INSPECTION INFORMATION): EVENT: PROCESSOR FRONT PANEL BUTTON AUTOMATICALLY TURN OFF.: DUE TO FAULTY RECEPTACLE UNIT. PAE JUDGMENT: PAE YES. INVESTIGATION REQUIRED: NECESSARY. TIER CLASSIFICATION: 2. UNIVERSAL FAILURE CODE: CVP3132Y. OTHER EVENTS ARE TIER 3 (PAE NO) AND DO NOT REQUIRE INVESTIGATION. (INSPECTION INFORMATION): EVENT: CVP1142F-FLICKERING IN IMAGE DUE TO FAULTY RECEPTACLE UNIT. EVENT: CVP3112A-FRONT PANEL SWITCHES ARE DEFORMED. CONTAINMENT MEASURES: NO (AT THIS TIME, NO FACTS HAVE BEEN CONFIRMED TO SUGGEST THAT THE REPORTED EVENT MAY EXPAND.).
IT WAS REPORTED THAT THE VIDEO SYSTEM CENTER PROCESSOR FRONT PANEL BUTTON GETS TURNED OFF AUTOMATICALLY. THE EVENT OCCURRED DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 544397 | VIDEO SYSTEM CENTER | VIDEO SYSTEM CENTER | FDF | SHIRAKAWA OLYMPUS CO., LTD. | CV-V1 | 04953170356384 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |