LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES
MDR report key: 1964693
·
Received December 23, 2010
Report
- Report Number
- 3015876-2010-01420
- Event Type
- Malfunction
- Date Received
- December 23, 2010
- Date of Event
- November 30, 2010
- Report Date
- November 30, 2010
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- MKJ
- PMA / PMN Number
- K991910
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED ENERGY CHARGE FAILURE. PHYSIO IS IN THE PROCESS OF REPAIRING THE DEVICE AND CONTINUES TO INVESTIGATE THE REPORTED EVENT. PHYSIO WILL SUBMIT A SUPPLEMENTAL REPORT ON THIS EVENT TO THE FDA AS PROVIDED BY 21 CFR 803.56.
Description of Event or Problem · 1
A THIRD PARTY BIOMED REPORTED THAT THE DEVICE WOULD INTERMITTENTLY FAIL TO CHARGE DEFIBRILLATION ENERGY AND ILLUMINATED THE SERVICE ICON. THERE WAS NO KNOWN PATIENT USE ASSOCIATED WITH THE EVENT.