FDA Adverse Event Malfunction Summary report: N

LIFEPAK 500 AUTOMATED EXTERNAL DEFIBRILLATOR

MDR report key: 1964662 · Received December 23, 2010

Report

Report Number
3015876-2010-01418
Event Type
Malfunction
Date Received
December 23, 2010
Date of Event
December 1, 2010
Report Date
December 1, 2010
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
MKJ
PMA / PMN Number
K955854
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. THE BATTERY WAS REPLACED AND THEN PROPER DEVICE OPERATION WAS OBSERVED THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHEN THE DEVICE WAS POWERED ON FOR TESTING, IT WOULD START REBOOTING ITSELF UNTIL THE BATTERY WAS REMOVED. THE REPORTED FAILURE WAS INTERMITTENT THROUGHOUT THE TESTING PROCESS. THERE WAS NO PT USE ASSOCIATED WITH THE REPORTED FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 500 AUTOMATED EXTERNAL DEFIBRILLATOR MKJ PHYSIO-CONTROL, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA