FDA Adverse Event Malfunction Summary report: N

REFURBED COLLEAGUE CX TRIPLE CCHANNEL VOLUMETRIC INFUSION PU

MDR report key: 1964621 · Received January 20, 2011

Report

Report Number
6000001-2011-00351
Event Type
Malfunction
Date Received
January 20, 2011
Date of Event
January 12, 2011
Report Date
January 13, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE CONDITION OF A COLLEAGUE INFUSION PUMP WITH AN OCCLUSION ALARM ON CHANNEL A WAS CONFIRMED AND DUPLICATED DURING PRODUCT EVALUATION. THE ASSIGNABLE CAUSE WAS DETERMINED TO BE A DISTAL 25 PSI FAILURE. THE PUMP HEAD MODULE CHANNEL A WAS REPLACED TO FIX THE REPORTED CONDITION. THIS DEVICE IS AN UNREMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.04.00. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED CONDITION. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). PER THE CUSTOMER, THE DEVICE IS AVAILABLE FOR EVALUATION; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. SHOULD THE DEVICE OR ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH AN OCCLUSION ALARM ON CHANNEL A, WHICH WAS DISCOVERED DURING BIO-MEDICAL TESTING. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS DEVICE IS CURRENTLY UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 REFURBED COLLEAGUE CX TRIPLE CCHANNEL VOLUMETRIC INFUSION PU PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1