FDA Adverse Event
Malfunction
Summary report: N
AFFINITY 4
MDR report key: 1964529
·
Received December 22, 2010
Report
- Report Number
- 1824206-2010-11890
- Event Type
- Malfunction
- Date Received
- December 22, 2010
- Date of Event
- November 30, 2010
- Report Date
- December 10, 2010
- Manufacturer
- HILL-ROM, INC.
- Product Code
- HDD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND THAT THE BRAKE DETENT ASSEMBLY WAS BROKEN, THE BRAKE/STEER CASTER HAD BEEN MOVED TO THE WRONG POSITION AND TWO OF THE BRAKE CASTERS WERE DAMAGED AND NON FUNCTIONAL. THE TECHNICIAN MOVED THE BRAKE/STEER CASTER TO THE CORRECT POSITION, REPLACED THE BRAKE DETENT ASSEMBLY, REPLACED THE TWO DAMAGED BRAKE CASTERS AND ADJUSTED THE BRAKE TO REPAIR THE BED.
Description of Event or Problem · 1
INFO REC'D INDICATES THE BRAKES ARE NOT WORKING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AFFINITY 4 | BIRTHING BED | HDD | HILL-ROM, INC. | 3700B |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |