FDA Adverse Event Malfunction Summary report: N

ADVANTA

MDR report key: 1964526 · Received December 22, 2010

Report

Report Number
1824206-2010-11896
Event Type
Malfunction
Date Received
December 22, 2010
Date of Event
December 5, 2010
Report Date
December 5, 2010
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN REPLACED THE POWER SUPPLY BOARD AND THE BED FUNCTIONED PROPERLY.

Description of Event or Problem · 1

TECHNICIAN ALLEGED THAT THIS BED HAD NO BED FUNCTIONS. TECHNICIAN DISCOVERED THAT THE POWER SUPPLY BOARD HAD FLUID INGRESS. NO ALLEGED INJURIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANTA A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1604

Patients

Seq Age Sex Outcome Treatment
1