FDA Adverse Event Malfunction Summary report: N

634 AFFINITY

MDR report key: 1964507 · Received December 22, 2010

Report

Report Number
1824206-2010-11878
Event Type
Malfunction
Date Received
December 22, 2010
Date of Event
December 3, 2010
Report Date
December 3, 2010
Manufacturer
HILL-ROM INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ACCOUNT FOUND THE TRENDELENBURG LINKAGE IS BENT ON THE PATIENT RIGHT SIDE FROM THE TRENDELENBURG BOLT BACKED OUT. THE ACCOUNT REPLACED THE TRENDELENBURG BRACKET AND BOLT ON THE PATIENT RIGHT SIDE TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

ACCOUNT ALLEGED THAT THE BED IS NOT GOING INTO TRENDELENBURG. NO INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 634 AFFINITY BIRTHING BED HDD HILL-ROM INC. 3700

Patients

Seq Age Sex Outcome Treatment
1