FDA Adverse Event
Malfunction
Summary report: N
634 AFFINITY
MDR report key: 1964507
·
Received December 22, 2010
Report
- Report Number
- 1824206-2010-11878
- Event Type
- Malfunction
- Date Received
- December 22, 2010
- Date of Event
- December 3, 2010
- Report Date
- December 3, 2010
- Manufacturer
- HILL-ROM INC.
- Product Code
- HDD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ACCOUNT FOUND THE TRENDELENBURG LINKAGE IS BENT ON THE PATIENT RIGHT SIDE FROM THE TRENDELENBURG BOLT BACKED OUT. THE ACCOUNT REPLACED THE TRENDELENBURG BRACKET AND BOLT ON THE PATIENT RIGHT SIDE TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
ACCOUNT ALLEGED THAT THE BED IS NOT GOING INTO TRENDELENBURG. NO INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 634 AFFINITY | BIRTHING BED | HDD | HILL-ROM INC. | 3700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |