FDA Adverse Event Malfunction Summary report: N

MINICAP EXTEND LIFE PD TRANSFSET W/TWIST CLAMP

MDR report key: 1964501 · Received January 20, 2011

Report

Report Number
1423500-2011-00814
Event Type
Malfunction
Date Received
January 20, 2011
Date of Event
December 28, 2010
Report Date
December 31, 2010
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
PMA / PMN Number
K882498
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE HAS BEEN REPORTED TO BE AVAILABLE FOR EVALUATION AND HAS BEEN REQUESTED. HOWEVER, THE SAMPLE HAS NOT BEEN RECEIVED FOR EVALUATION AS OF YET. IF THE SAMPLE IS RECEIVED, A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION.

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR A LEAK WAS CONFIRMED IN THE LAB. THE RESULTS OF A SAMPLE EVALUATION REVEALED THAT BOTH OF THE OCCLUDER FEET WERE BROKEN AND CONTAINED WHAT APPEARED TO BE DIRT AND RESIDUE. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT. A ROOT CAUSE WAS NOT IDENTIFIED DUE TO INSUFFICIENT INFORMATION. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

A PATINET CONTACTED BAXTER (B)(4) TO REPORT THAT AFTER ONE MONTH OF USE LIQUID CAN NOT BE STOPPED EVEN IF CLOSING THE CLAMP ON THE TRANSFER SET. THERE WAS NO USE OF ADDIONAL DISINFECTANT. THERE WAS PATIENT INVOLVEMENT, BUT NO PAITENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.