FDA Adverse Event Injury Summary report: N

SELECT SECURE

MDR report key: 1964460 · Received January 20, 2011

Report

Report Number
2649622-2011-00080
Event Type
Injury
Date Received
January 20, 2011
Date of Event
December 22, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVN
PMA / PMN Number
P030036
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT ALL CONDUCTORS WERE DISTORTED, THE OUTER INSULATION WAS MELTED, THE OUTER INSULATION WAS PARTIALLY OR FULLY COVERING THE TIP ELECTRODE, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THERE WAS APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THE RIGHT VENTRICULAR LEAD DISLODGED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SELECT SECURE IMPLANTABLE PACING LEAD NVN MEDTRONIC PUERTO RICO, INC. 3830 ASKU

Patients

Seq Age Sex Outcome Treatment
1 12 YR Hospitalization| R ADDR01 IMPLANTABLE PULSE GENERATOR| 3830 IMPLANTABLE PACING LEAD