FDA Adverse Event
Injury
Summary report: N
SELECT SECURE
MDR report key: 1964460
·
Received January 20, 2011
Report
- Report Number
- 2649622-2011-00080
- Event Type
- Injury
- Date Received
- January 20, 2011
- Date of Event
- December 22, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- NVN
- PMA / PMN Number
- P030036
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT ALL CONDUCTORS WERE DISTORTED, THE OUTER INSULATION WAS MELTED, THE OUTER INSULATION WAS PARTIALLY OR FULLY COVERING THE TIP ELECTRODE, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THERE WAS APPARENT EXPLANT DAMAGE.
Description of Event or Problem · 1
IT WAS REPORTED THE RIGHT VENTRICULAR LEAD DISLODGED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SELECT SECURE | IMPLANTABLE PACING LEAD | NVN | MEDTRONIC PUERTO RICO, INC. | 3830 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 12 YR | Hospitalization| R | ADDR01 IMPLANTABLE PULSE GENERATOR| 3830 IMPLANTABLE PACING LEAD |