MEDISENSE OPTIUM
Report
- Report Number
- 2954323-2011-00753
- Event Type
- Malfunction
- Date Received
- January 19, 2011
- Date of Event
- December 23, 2010
- Report Date
- January 19, 2011
- Product Code
- NBW
- Removal / Correction Number
- ADC FA1197-2010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
RETAIN SAMPLES OF PRECISION TEST STRIP LOTS MANUFACTURED WITH HOT-MELT GLUE BATCHES EXHIBITED LOWER THAN EXPECTED READINGS ON CONTINUOUS STORAGE AFTER NINE MONTHS AT 30C DURING ROUTINE STABILITY PERFORMANCE TESTING. THIS STABILITY ISSUE COULD LEAD TO THE GENERATION OF INCORRECTLY DEPRESSED BLOOD GLUCOSE RESULTS AND THESE ERRONEOUS RESULTS MAY BE IN THE C, D OR E ZONES OF PARKES ERROR GRID, AND AS SUCH, HAVE THE POTENTIAL TO BE CLINICALLY SIGNIFICANT. THE PRIMARY CAUSE HAS BEEN IDENTIFIED TO BE BATCH TO BATCH GLUE VARIABILITY. THE FDA HAS BEEN INFORMED OF THE FIELD ACTION ((B)(4)). (B)(4).
IN USING AN AFFECTED TEST STRIP RELATED TO AN ON-GOING FIELD ACTION FOR ABBOTT'S PRECISION FAMILY TEST STRIPS, THE CUSTOMER REPORTED EITHER LONGER BLOOD FILL TIME OR READINGS ISSUES. THE CUSTOMER REPORTED GETTING LOWER THAN FEELS READINGS AND EXPERIENCING VERTIGO, TREMULOUSNESS, DISORIENTATION, FATIGUE, BLURRED VISION, DIFFICULTY BREATHING AND NAUSEA. THE CUSTOMER REPORTEDLY WAS SEEN BY A DOCTOR AND DIAGNOSED WITH HYPOGLYCEMIA. IT WAS FURTHER STATED HE CONSULTED TWO DOCTORS AND ONE PUT HIM ON SUPPLEMENTS AND THE OTHER HAD HIM ON SMALL MEALS EVERY 2 HOURS. HE ALSO SELF-TREATED BY DRINKING HIGH-PROTEIN DRINK OR EATING FOOD TO ALLEVIATE THE SYMPTOMS. THERE WAS NO REPORT OF DEATH OR SERIOUS INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MEDISENSE OPTIUM | BLOOD GLUCOSE MONITORING SYSTEM | NBW | 46332 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |