FDA Adverse Event Malfunction Summary report: N

AUTOSOFT XC

MDR report key: 19644082 · Received July 1, 2024

Report

Report Number
3003442380-2024-11028
Event Type
Malfunction
Date Received
July 1, 2024
Date of Event
May 20, 2024
Report Date
October 30, 2024
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244016927
PMA / PMN Number
K032854
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: THE INVESTIGATION ASSOCIATED WITH RELATED EVENT DATABASE 1900839 HAS BEEN APPROVED AND IS COMPLETE. NO ADDITIONAL ACTION IS REQUIRED, AND THIS COMPLAINT WILL BE CLOSED. THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKETING SURVEILLANCE (PMS) PRODUCT TRENDS AND MALFUNCTION ACCORDING TO THE MARKET QUALITY REVIEW (OMQR) PROCEDURE. THE IDENTIFIED FOR MALFUNCTION CODE, SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE, IS NOT ASSOCIATED WITH A DESIGN OR MANUFACTURING-RELATED COMPLAINT ISSUE. THEREFORE, A DETAILED INVESTIGATION OR INSPECTION OF REFERENCE SAMPLES IS NOT REQUIRED. BATCH REVIEW LOT 6004680 WAS MANUFACTURED ON 05-DEC-2023, IN LINE 9, WITH A TOTAL OF 3,600. THE BATCH RECORD WAS REVIEWED TO VERIFY IF ALL THE APPLICABLE PROCEDURES WERE FOLLOWED, AND NO ISSUES WERE FOUND. REVIEW OF THE BATCH RECORD SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE MANUFACTURING PROCESS AND FINAL PRODUCT RELEASE HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DISCREPANCY RELATED TO THIS ISSUE WAS FOUND WITHIN THE DOCUMENTATION. CONCLUSION SUMMARY OF THE RELATED EVENT. DUE TO THE FOLLOWING BATCH RECORD REVIEW YIELDING NO DISCREPANCIES AND NO NON-CONFORMANCE (NC) WAS GENERATED DURING PRODUCTION. DURING MANUFACTURING OF THE CANNULA PART, THE SOFT CANNULA IS KEPT STRAIGHT BY THE INTRODUCER NEEDLE, NO SAMPLES WERE RETURNED FOR INSPECTION. HOWEVER, DURING USE IN GENERAL AND SPECIFICALLY DURING INSERTION MAY ACCIDENTALLY KINK THE SOFT CANNULA. NO FURTHER ACTION IS REQUIRED FOR THIS COMPLAINT, IF USED/UNUSED SAMPLES ARE RECEIVED, APPROPRIATE ACTIONS WILL BE TAKEN.

Description of Event or Problem · 0

UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT, PATIENT FACED INFUSION SET CANNULA KINNKED EVENT ON 20-MAY-2024. THE EVENT OCCURRED WITHIN 3 HOURS OF INSERTION. THE INSERTION SITE WAS AT ABDOMEN. THE BLOOD GLUCOSE LEVEL OF THE PATIENT WAS 200-363 MG/DL. UNOMEDICAL DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION AVAILABLE

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1256468 AUTOSOFT XC UNO INSET I 60/6 GREY TCAP 10PK INT FPA UNOMEDICAL A/S 1001680 6004680 05705244016927

Patients

Seq Age Sex Outcome Treatment
1 64 YR Female