FDA Adverse Event Malfunction Summary report: N

UNICEL® DXI 800 ACCESS® IMMUNOASSAY ANALYZER

MDR report key: 1964377 · Received January 19, 2011

Report

Report Number
2122870-2011-00075
Event Type
Malfunction
Date Received
January 19, 2011
Date of Event
November 29, 2010
Report Date
December 21, 2010
Manufacturer
BECKMAN COULTER, INC.
Product Code
JJE
PMA / PMN Number
K023764
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SAMPLE WAS COLLECTED IN A GREINER LITHIUM HEPARIN PLASMA TUBE. THE CUSTOMER CENTRIFUGES SAMPLES FOR 15 MINUTES AT 2200 RPM. HOWEVER, SOME SAMPLES ARE CENTRIFUGED FOR 6 MINUTES AT 2000 RPM. QC HAS BEEN WITHIN THE ESTABLISHED RANGES. SERVICE WAS NOT DISPATCHED FOR THIS EVENT AS THE CUSTOMER IS NOT QUESTIONING INSTRUMENT PERFORMANCE AT THIS TIME. A CLEAR ROOT CAUSE FOR THE EVENT HAS NOT BEEN DETERMINED TO DATE.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO ERRATIC TOTAL BETA HCG (TBHCG) RESULTS IN DIFFERENT GESTATIONAL AGE CATEGORIES GENERATED BY UNICEL DXI 800 ACCESS IMMUNOASSAY ANALYZER FOR ONE PATIENT. THE RESULT WAS NOT REPORTED OUT OF THE LABORATORY. SUBSEQUENT TESTING PRODUCED REPRODUCIBLE RESULTS. THE CUSTOMER DID NOT RECEIVE ANY REPORT OF PATIENT INJURY REQUIRING MEDICAL INTERVENTION OR CHANGE TO PATIENT TREATMENT ATTRIBUTED OR CONNECTED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXI 800 ACCESS® IMMUNOASSAY ANALYZER DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER JJE BECKMAN COULTER, INC. DXI 800 NA

Patients

Seq Age Sex Outcome Treatment
1